23
Symbol on select
or switch
Selected mode
Standby mode
►
NOTICE! Only use the vacuum pump when it is installed on the adapter plate.
Use the selector switch to choose the desired mode. The vacuum pump reports certain oper
ating states with signals.
Light signal
Audible signal
Status
LEDs 1 and 2 flash white
Signal tone rising (1x)
Restart of the device
LEDs 1 and 2 illuminated red
Signal tone falling (1x)
Device switching to standby mode
–
Click sound
Selector switch operated, mode
change
LEDs 1 and 2 flash blue
–
Pump running (forwards or in
reverse)
LEDs 1 and 2 flash green, every 10 s
Device is powered up and working
without leaks
LEDs 1 and 2 flash alternating blue
and red, every 10 s
–
Leak warning (see information)
LEDs 1 and 2 flash red
Signal tone, like siren (1x)
System error
Leak warning information
The leak warning indicates that the desired vacuum cannot be achieved or maintained. The vacu
um pump continues running. The leak warning is output under the following conditions:
•
The vacuum pump operates continuously for more than 3 minutes
•
Less than 2 minutes elapse between several pump intervals
The leak warning deactivates itself automatically when there is no problem for 10 minutes. It can
also be deactivated by briefly activating the vacuum pump standby mode.
6.3 Flushing the Vacuum Pump
A reduced vacuum indicates contamination. In case of contamination, the system has to be
flushed to clean it. This flushes the vacuum pump and the filter in the adapter plate. The reverse
mode of the vacuum pump is used for flushing. The flushing process can also help if the system is
plugged.
CAUTION
Improper cleaning
Skin irritation, formation of eczema or infections due to contamination with germs
►
Clean the product regularly.
>
Required materials
:
Vessel with 30 ml to 60 ml distilled water, one collecting vessel
Preparations required:
Vacuum pump installed on adapter plate and in standby mode, prosthesis taken off the leg
1) Dip the outlet hose into the vessel with distilled water.
2) Hold the socket in such a way that the water runs into the collecting vessel during flushing.
3) Set the selector switch to reverse mode.
4) Allow the vacuum pump to run until the water has been pumped through the entire system.
5) Allow the pump to dry run for 15 s in order to pump out any remaining water.
6) Set the selector switch to standby mode.
7) Clean the interior surface of the socket.
7 Maintenance
►
A visual inspection and functional test of the prosthetic components should be performed
after the first 30 days of use.
Summary of Contents for Harmony E2 4R152
Page 2: ...1 2 3 4 5 6 2...
Page 3: ...7 8 9 10 11 12 3...
Page 4: ...13 14 15 16 17 18 4...
Page 5: ...19 20 21 22 23 24 5...
Page 119: ...119 3 3 1 3 2 118...
Page 122: ...122 5 1 2 2 3 4 6 1 8 2 9 3 USB 10 1 2 3 4 11 5 6 7 1 2 1 1 1 2 3 1 2 1 2 1 2...
Page 123: ...123 1 2 5 2 Harmony 5 2 1 5R2 TEC Plate 4A94 3 6 12 M12x1 5 5R2 TEC Plate 1 3 13 2 14 3 15...
Page 125: ...125 5 3 2 1 5 2 3 4 6 6 1 1 2 20 3 21 6 2 4 1 4...
Page 126: ...126 1 2 1 1 2 1 1 2 1 2 10 1 2 10 1 2 1 3 2 10 6 3 30 60 1 2 3 4 5 15...
Page 129: ...129 3 3 1 3 2 128 1...
Page 135: ...135 6 2 4 1 4 LED1 2 1 LED1 2 1 LED1 2 LED1 2 10 LED1 2 10 LED1 2 1 3 2 10 6 3...
Page 137: ...137...
Page 138: ...138...
Page 139: ...139...