20
5) After the charging process is complete, disconnect the plug from the charging receptacle and
close the charging receptacle cover.
6) Disconnect the battery charger plug from the outlet.
7)
If necessary:
Connect the outlet hose and attach the vacuum pump to the adapter plate.
Light Signal
Audible Signal
Status
LEDs 1 and 2 flash alternating yellow
(until automatic shutdown)
Signal tone ringing (1 x)
Batteries almost drained, remaining
capacity ≤ 1 h
LEDs 1 and 2 flash alternating red (until
automatic shutdown)
Signal tone ringing (3 x)
Batteries drained, device shuts
down
LEDs 1 and 2 flash alternating green
–
Charger plugged in, connection
established
LEDs 1 and 2 flash alternating green,
brightness dimmed
–
Batteries are being charged
LEDs 1 and 2 illuminated green, bright
ness dimmed
–
Batteries fully charged
LEDs 1 and 2 illuminated turquoise,
brightness dimmed
–
Batteries fully charged, reduced
battery performance; we recom
mend replacing the batteries
5.2 Installing the Adapter Plate
The adapter plate for connecting the vacuum pump is installed between the prosthetic socket and
the distal components of the modular prosthesis. It is mounted together with the socket adapter.
The adapter plate has two connectors for generating a vacuum in the socket and one connector
for connecting to the vacuum pump. The adapter plate is additionally equipped with a PE sinter
filter to protect the vacuum pump against dirt, and a non-return valve that maintains the vacuum
when the pump is disconnected.
CAUTION
Using a wrong prosthetic socket
Contusions and pressure sores due to wrong socket fit
►
Use the vacuum pump only in combination with a prosthetic socket made with Harmony
technology.
5.2.1 Preparing the Adapter Plate
A vacuum can be generated in the socket either via the proximal or the lateral connector of the
adapter plate.
INFORMATION
Using the proximal connector of the adapter plate is recommended only when using a 5R2 lam
ination disc or the 4A94=3 TEC plate (with TPU insert respectively). The lateral connector for
the intake hose can be recognised by the internal thread for the hose connector.
Using the proximally located connector
INFORMATION
To ensure an optimum sealed connection to the socket, the following approach is recommen
ded:
►
Drill the distal socket bore with a
6 mm
bit.
►
Apply PU adhesive inside the bore and insert the included hose piece (see fig. 12).
►
Remove excess adhesive and allow the adhesive to set.
►
Trim the hose piece flush on both ends.
Summary of Contents for Harmony E2 4R152
Page 2: ...1 2 3 4 5 6 2...
Page 3: ...7 8 9 10 11 12 3...
Page 4: ...13 14 15 16 17 18 4...
Page 5: ...19 20 21 22 23 24 5...
Page 119: ...119 3 3 1 3 2 118...
Page 122: ...122 5 1 2 2 3 4 6 1 8 2 9 3 USB 10 1 2 3 4 11 5 6 7 1 2 1 1 1 2 3 1 2 1 2 1 2...
Page 123: ...123 1 2 5 2 Harmony 5 2 1 5R2 TEC Plate 4A94 3 6 12 M12x1 5 5R2 TEC Plate 1 3 13 2 14 3 15...
Page 125: ...125 5 3 2 1 5 2 3 4 6 6 1 1 2 20 3 21 6 2 4 1 4...
Page 126: ...126 1 2 1 1 2 1 1 2 1 2 10 1 2 10 1 2 1 3 2 10 6 3 30 60 1 2 3 4 5 15...
Page 129: ...129 3 3 1 3 2 128 1...
Page 135: ...135 6 2 4 1 4 LED1 2 1 LED1 2 1 LED1 2 LED1 2 10 LED1 2 10 LED1 2 1 3 2 10 6 3...
Page 137: ...137...
Page 138: ...138...
Page 139: ...139...