Fasten the unit from the outside with PVC adhesive tape. Seal off the unit as much as possible
by pulling the tape tight.
NOTICE
Risk of damage to the receptacle.
Sealing resin inside the receptacle can cause damage. Resin
dissolves the material of the receptacle. Do not let resin penetrate into the receptacle.
INFORMATION
Ensure that the cable connection points in the right direction: Knee joint cable toward the electronics,
electronics cable toward the battery!
Turn the casing so that you can see the receptacle from the inside. Mix a small amount of sealing
resin with a sufficient amount of hardener. Do not add too much hardener, since this could cause
the resin to form bubbles during hardening and make the connection brittle.
Bond the receptacle to the casing. In order to do so, let the sealing resin flow gently and slowly
around the receptacle into the recess (Figure 36).
36
Then glue in the battery receptacle. Follow the above-mentioned instructions for gluing the bat-
tery receptacle. Put the casing aside for hardening.
Finish the orthosis in the usual manner: Mount the knee and the ankle joint if applicable, and
attach inner pads and closures.