2
8.3 Programming Tools
WARNING
Risk of injury as a result of incorrect programming.
Programming may only be conducted by specialist
staff with in-depth knowledge of the VR2 control. Incorrect programming could result in an unsafe set-
up of a wheelchair for a user. Otto Bock or the manufacturer of the control accept no responsibility for
losses of any kind if the programming of the control system is altered from the factory pre-set values.
8.3.1 Hand Programming Device
INfORMATION
If you have a PP1a, read the PP1a user guide before you use it.
If you re-program your control system, make sure that you observe any restrictions given in your wheel-
chair user manual.
Document any changes you make for future reference.
The PP1a hand programming device gives access to programmable parameters in order to
adjust the wheelchair to individual patient requirements and to read the Fault Log and Timer
(see Section 8.). To alter the program settings, the device is connected to the control system
with a plug.
For details of how to use the hand programming device refer to the "Programming and Diagnos-
tics” guide supplied with the PP1a hand programming device.
8.3.2 PC Programming Device
The PC programming device gives the same access level as the PP1a hand programming device.
For details of how to use these software packages with the VR2 control, refer to the documenta-
tion supplied with the software.