74
1) Connect the inductive charger to the receiver of the charging unit
on the rear of the product. The charger is held in place by a mag
net.
→
The LED ring on the rear of the charger pulsates purple
(4-second cycle).
→
If the LED ring lights up in a different colour, this indicates an
error (see Page 99).
2) The charging process starts.
→
Once the product battery is fully charged, all LEDs on the side
of the battery charger light up.
3) After the charging process is complete, hold the prosthesis still and
remove the inductive charger from the receiver.
→
A self-test is performed. The joint is operational only after cor
responding feedback (see Page 102).
7.3 Display of the current charge level
7.3.1 Display of battery charge level without additional devices
INFORMATION
The charge level cannot be displayed during the charging process, e.g. by turning the prosthes
is over. The product is in charging mode.
1) Turn the prosthesis by 180° (the sole of the foot must face up).
2) Hold still for 2 seconds and wait for beep signals.
Beep signal
Battery charge level
5x short
more than 80%
4x short
60% to 80%
3x short
40% to 60%
2x short
20% to 40%
1x short
less than 20%
INFORMATION
If the
Volume
parameter is set to '0' in the Cockpit app (see Page 83) or if mute mode (silent
mode) is activated, there are no beep signals.
7.3.2 Display of the current charge level using the Cockpit app
Once the Cockpit app has been started, the current charge level is displayed in the bottom line of
the screen: