21
VIII. SPECIFICATIONS
Temperature range:
12
°
to 50
°
C
Cool stimulus:
12
°
to 37
°
C (Recommended setting: 24
°
C)
Warm stimulus:
37
°
to 50
°
C (Recommended setting: 50
°
C)
Temperature readout:
Digital
Time to temperature:
< 60 seconds
Temperature accuracy:
+/-
0.3
°
C
Time range:
1 to 99 seconds (Recommended setting: 60 seconds)
Flow rate range:
4 - 10 LPM (Recommended setting: 8 LPM)
Air supply:
Internally mounted compressor
Remote control capability:
Standard for computer-based systems.
Timeout indication:
Beeper
Input power:
120 VAC @ 60 Hz or 240 VAC @ 50 Hz
Current:
5A @ 120 VAC, 3A @ 240 VAC
Safety:
Primary temperature read-out. Separate over-temperature
cut-off.
Stimulus timing activation: Switch
on delivery head or footswitch.
Size (W X D X H):
14 x 17.5 x 7.8 inches
35 x 44 x 20 centimeters
Weight:
43 lbs.
20 kg.
Operating Environment
Temperature:
+15
°
C to +35
°
C (+59
°
F to +95
°
F)
Relative Humidity:
30 to 90%, non-condensing
Air Pressure:
600 hPa to 1060 hPa
Operations at temperatures below -20
°
C or above +60
°