1. Computer Connection
The stimulator cable connector goes to the computer connection on the
rear panel of the stimulator. Through this connection to the computer,
the computer is able to select stimulus temperature through the software.
NOTE: This selection can be made on the unit's front panel (the orange
and blue temperature select controls).
2.
Remote Footswitch Connection
A footswitch can be connected to the remote control connector at the
rear of the stimulator. This would be used to activate the flow timer. If
a computer is connected and a caloric test has been selected in the
software, the first depression of the footswitch will initiate the flow
timer and the recording; subsequent depressions will center the tracing
(ENG/VNG) or start the video recording (VNG only). If the computer is
turned on but the software is not running, flow time may be activated by
holding down the flow time activation switch or the footswitch for 5
seconds. The functions of the footswitch operate in parallel with the
flow time activation switch on the delivery head.