• Is the patient well at rest and is there no muscle tension? Note that electric mus-
cle artifacts may arise from non-moving but tensed muscles (isometric contrac-
tion). Please check especially the nape of the neck for this condition.
• Are there any sources of electric noise in the environment? Especially motor-
driven devices such as pumps can cause high noise which is sometimes accepted
by the device but can be too high, if for instance the response has a low ampli-
tude. It may be worth while to relocate to another room for testing.
• Is there much acoustic noise in the environment? A high level of noise might
mask the stimulus.
Incomplete test
5.6
Test view
A patient screened by AccuScreen is either given a
PASS (Test passed)
or a
REFER
(Test referred)
.
• If the result is a
PASS (Test passed)
, no further action is required.
• If the result is a
REFER (Test referred)
, it is recommended that you retest the
patient. If the patient is given yet another
REFER (Test referred)
, this patient
should be referred for further testing.
5.6.1
Test View
You can view a patient’s tests:
• from the
Home
screen
• from the
Patient Details
screen
If the test was stopped, an error message is shown.
Summary of Contents for AccuScreen ABR
Page 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Page 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Page 66: ...66 Otometrics Printing Printouts from label printer ...
Page 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Page 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Page 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...