Hygiene and maintenance
Sterilization
Measuring probe with cable / Universal support
W&H recommends sterilization according to EN 13060, EN 285.
> Read the notes, follow the instructions and heed the warnings provided by the manufacturers of steam sterilizers.
> The program selected must be suitable for the measuring probe with cable and the universal support.
Recommended sterilization cycles
> Steam sterilization (type B, S)
> Sterilization time at 3 minutes at 134°C (273°F), 4 minutes at 132°C (270°F)
> Maximum sterilization temperature 135°C (275°F)
Evidence of the measuring probe with cable and the universal support basic suitability for effective sterilization was provided
by an independent test laboratory using the LISA 517 B17L steam sterilizer (W&H Sterilization S.r.l., Brusaporto (BG)) and the
Systec VE-150 steam sterilizer (Systec).
> »Dynamic-air-removal prevacuum cycle« (type B): temperature 134°C (273°F) – 3 minutes*
temperature 132°C (270°F) – 4 minutes*
> »Steam-flush pressure-pulse cycle« (type S): temperature 134°C (273°F) – 3 minutes*
* EN 13060, EN 285, ISO 17665