Figure 31:
Dose Rate Settings
This setting is designed for your personal safety. The alarm is intended to let you know you are
exposed to dangerous radiation and may be accumulating a significant radiation dose. If you set the
alarm levels or warning thresholds too high, this may pose a serious health risk to you.
3.7 Detect and ID Settings
3.7.1 Setup of Statistical Characteristics of the Detect Mode
The functionality in
DETECT MODE
depends on a statistical assessment of the count rate time series.
You may change the sensitivity within the
DETECT MODE
by adjusting the settings in this tab.
3.7.2 Definition of the Easy-ID Time Preset
The
EASY-ID MODE
performs an identification based on a fixed ID Time in seconds. You can change this
time preset by entering a different number in the associated tab. The default setting for the ID Time is
60 seconds. Figure
, p.
shows the
DETECT/ID SETTINGS
screen.
3.8 Editing the Nuclide Library
3.8.1 The Nuclide Library
The RADEAGLET has a radionuclide library. It identifies those nuclides that are in the list and activated.
Upon entering the
NUCLIDE SETTINGS
screen, you will see a table with the nuclide short name as first
© innoRIID GmbH
RADEAGLET User Manual• Software 2.1.16 • Document 1.32o • 2017-12-15
Page 39 of 60