APPENDIX C. FIRMWARE COMMANDS
47
C.3.11 Set Real Time Data Recording Mode On/Off
NOTE
Sends a 20 byte data packet every 250 ms.
Function
Code
Description
Char + ‘|’
0x7C
Header 1
Char + ‘&’
0x26
Header 2
CTRL + ‘T’
0x14
Set the Real Time Data Reporting Mode ON/OFF
C.3.12 Operational Notes
1)
All computer commands sent to the Counter are accepted only when you are in the Stop Mode; except
for Stop, Send Parameters, and Real Time Data Mode commands, which can be sent anytime.
2)
Sending any command other than Send Parameters will take the counter out of Local mode and place it
into Remote mode.
3)
When the Counter is in Remote mode it will not respond to any front panel control buttons or switch
settings, except for Display mode and Display OVF switches.
4)
To return to Local mode, cycle the Counter power switch off then on.
CAUTION:
All data will be lost.