ORTEC
®
EASY-MCA™ Digital Gamma-Ray Spectrometer User’s Manual
931044C / 1014
24
131
Command execution error
132
Invalid Command
For macro code 129 (syntax error) or 131 (execution error), the following apply:
1
Invalid Verb
2
Invalid Noun
4
Invalid Modifier
128
Invalid first parameter
129
Invalid second parameter
130
Invalid third parameter
131
Invalid fourth parameter
132
Invalid number of parameters
133
Invalid command
134
Response buffer too small
135
Not applicable while active
136
Invalid command in this mode
137
Hardware error
138
Requested data not found
Micro error codes:
0
Success
1
Input already started/stopped
2
Preset already exceeded
4
Input not started/stopped
64
Parameter was rounded (for decimal numbers)
128
No sample data available
4.2.1. Dollar Response Records
SHOW and STEP commands respond with a single dollar response record followed immediately
by a percent response record. All valid dollar response records for each command are listed in
the command dictionary.
The following list provides the general form of each dollar response record for the MCB API. In
this list, lower case letters represent numeric values. The letters “ccc” always represent an 8-bit