80
|
INSTRUCTIONS FOR USE
OM 602RS
ERROR
CAUSE
ELIMINATION
E. D. Un.
Number is too small (large negative) to be
displayed
change DP setting, channel constant setting
E. D. Ov.
Number is too large to be displayed
change DP setting, channel constant setting
E. T. Un..
Number is outside the table range
increase table values, change input setting (chan-
nel constant setting)
E. T. Ov.
Number is outside the table range
increase table values, change input setting (chan-
nel constant setting)
E. I. Un.
Input quantity is smaller than permitted input
quantity range
change input signal value or input (range) setting
E. I. Ov.
Input quantity is larger than permitted input
quantity range
change input signal value or input (range) setting
E. HW.
A part of the instrument does not work properly
send the instrument for repair
E. EE
Data in EEPROM corrupted
perform restoration of manufacture setting, upon
repeated error statement send instrument for repair
E.. DATA
Data in EEPROM outside the range
perform restoration of manufacture setting, upon
repeated error statement send instrument for repair
E. CLR.
Memory was empty
(presetting carried out)
upon repeated error statement send instrument for
repair, possible failure in calibration
ERROR STATEMENTS
9
Summary of Contents for OM 602RS
Page 1: ...OM 602RS 6 DIGIT PROGRAMABLE INSTRUMENT RS 232 RS 485 GUARANTEE Y E A R S ...
Page 7: ...INSTRUCTIONS FOR USE OM 602RS 7 ...
Page 31: ...INSTRUCTIONS FOR USE OM 602RS 31 5 light light SETTING ...
Page 39: ...INSTRUCTIONS FOR USE OM 602RS 39 5 light light SETTING ...
Page 41: ...INSTRUCTIONS FOR USE OM 602RS 41 6 profi profi SETTING ...
Page 57: ...INSTRUCTIONS FOR USE OM 602RS 57 6 profi profi SETTING ...
Page 71: ...INSTRUCTIONS FOR USE OM 602RS 71 6 profi profi SETTING ...
Page 75: ...INSTRUCTIONS FOR USE OM 602RS 75 6 profi profi SETTING ...
Page 83: ...INSTRUCTIONS FOR USE OM 602RS 83 11 TECHNICAL DATA PI Primary insulation DI Double insulation ...
Page 86: ...86 INSTRUCTIONS FOR USE OM 602RS NOTE ...