INSTRUCTIONS FOR USE
OM 602RS
|
49
6.1.2h
Selecting display mode in case of communication failure
6.1.2i
Setting the time constant for Timeout
TIMEOU.
MAX. 9F
CHANNE.
SERVIC.
OUTPUT.
CLEAR
EXT. IN.
CONFIG.
KEYS
INPUTS
COMMA.
BAUD
ADDR.
PROT.
MIN. 9N
MAX. 9N
MOD T.0.
NO
MIN. 9F
BLANK
FLASH
DASHES
DOT
DEF
MOD T.0.
Selecting display
mode in case of
communication failure
NO
No reaction
BLANK
Displey goes off
FLASH
Last displayed value
starts flashing
DASHES
Dash symbols displayed
DOT
Decimal point is
displayed
Item will not appear in “MASTER” protocol
!
TIMEOU.
MAX. 9F
MOD T.O.
CHANNE.
SERVIC.
OUTPUT.
CLEAR
EXT. IN.
CONFIG.
KEYS
INPUTS
COMMA.
BAUD
ADDR.
PROT.
MIN. 9N
MAX. 9N
1.0
MIN. 9F
DEF
TIMEOU.
Setting the time constant
for Timeout
- setting the time delay after which the
indication of interrupted communication
will appear on the display in the mode of
“Mod t.0.”
- range: 0…99,9 s
-
DEF
= 1.0 s
Item will not appear in “MASTER” protocol and when
“MOD t.0.” is disabled
!
6
profi
profi
SETTING
Summary of Contents for OM 602RS
Page 1: ...OM 602RS 6 DIGIT PROGRAMABLE INSTRUMENT RS 232 RS 485 GUARANTEE Y E A R S ...
Page 7: ...INSTRUCTIONS FOR USE OM 602RS 7 ...
Page 31: ...INSTRUCTIONS FOR USE OM 602RS 31 5 light light SETTING ...
Page 39: ...INSTRUCTIONS FOR USE OM 602RS 39 5 light light SETTING ...
Page 41: ...INSTRUCTIONS FOR USE OM 602RS 41 6 profi profi SETTING ...
Page 57: ...INSTRUCTIONS FOR USE OM 602RS 57 6 profi profi SETTING ...
Page 71: ...INSTRUCTIONS FOR USE OM 602RS 71 6 profi profi SETTING ...
Page 75: ...INSTRUCTIONS FOR USE OM 602RS 75 6 profi profi SETTING ...
Page 83: ...INSTRUCTIONS FOR USE OM 602RS 83 11 TECHNICAL DATA PI Primary insulation DI Double insulation ...
Page 86: ...86 INSTRUCTIONS FOR USE OM 602RS NOTE ...