optris CTratio
– E2013-04-A
29
P----
Setup of
Peak hold
. If the value is set to
0.0
the display will show
---
(function deactivated).
In this mode the sensor is waiting for descending signals. If the signal descends the
algorithm maintains the previous signal peak for the specified time.
After the hold time the signal will drop down to the second highest value or will descend by
1/8 of the difference between the previous peak and the minimum value during the hold time.
This value will be held again for the specified time. After this the signal will drop down with
slow time constant and will follow the current object temperature.
V----
Setup of
Valley hold
. If the value is set to
0.0
the display will show
---
(function deactivated).
In this mode the sensor waits for ascending signals. The definition of the algorithm is
according to the peak hold algorithm (inverted).
u 700.0
Setup of the
Lower limit of temperature range
. The minimum difference between lower
and upper limit is
20 K
. If you set the lower limit to a value
≥ upper limit the upper limit will be
adjusted to
[lower limit + 20 K]
automatically.
n 1800
Setup of the
Upper limit of the temperature range
. The minimum difference between
upper and lower limit is
20 K
. The upper limit can only be set to a value = lower limit + 20 K.
[ 0.00
Setup of the
Lower limit of the signal output
. This setting allows an assignment of a
certain signal output level to the lower limit of the temperature range. The adjustment range
corresponds to the selected output mode (e.g. 0-5 V).
] 5.00
Setup of the
Upper limit of the signal output
. This setting allows an assignment of a
certain signal output level to the upper limit of the temperature range. The adjustment range
corresponds to the selected output mode (e.g. 0-5 V).