optris CTratio
– E2013-04-A
28
S ON
Activating (
ON
) and Deactivating (
OFF
) of the integrated
Sighting Laser
By pressing
Up
or
Down
the laser will be switched on or off. If the laser is activated a
measurement cannot be done.
MV5
Selection of the
Output signal
. By pressing
Up
or
Down
the different output signals can be
selected
[► Outputs and Inputs].
R 1.000
Setup of
Slope
. Pressing
Up
increases the value,
Down
decreases the value (also valid
for all further functions). The slope is the quotient of the emissivities of both of the
overlapping wavelengths and therewith the deciding parameter for measurements in 2-color
mode.
[► Characteristic Slope Values]
E 0.970
Setup of
Emissivity
. The emissivity is a material constant factor to describe the ability of the
body to emit infrared energy. The emissivity only affects measurements in the 1-color mode.
[► Emissivity]
D 95.0
Setup of the allowed
Attenuation
. The temperature measurement will stop if the attenuation
exceeds this limit
[► Attenuation]
.
MD 2C
Setup of the
Main display
on the electronic box. By pressing
Up
or
Down
you can select if
the 2C temperature, 1C temperature or attenuation should be displayed (independent from
the selected output signal).
A 0.2
Setup of
Average time
. If the value is set to
0.0
the display will show
---
(function
deactivated). In this mode an arithmetic algorithm will be performed to smoothen the signal.
The set time is the time constant. This function can be combined with all other post
processing functions.