16
-
2.4 Measurement Specifications [2M models]
2ML
2MH
2MH1
Temperature range (scalable)
1 color: 250...1000 °C
2 color: 275...1000 °C
1 color: 375...1500 °C
2 color: 400...1500 °C
1 color: 5
00…3000 °C
2 color: 55
0…3000 °C
Spectral range
1,45 - 1,75 µm
Optical resolution
38:1
50:1
100:1
Vario focus
300 mm to infinity, infinitely adjustable
System accuracy
1), 2), 3)
±(0,5 % of r2 °C)
Repeatability
1), 2). 3)
±0,3 % of reading
Temperature resolution
0,1 K
Response time (90 % signal)
4)
1 ms…10 s
Emissivity
0,050…1,000 (adjustable via programming keys or analog input)
Slope
0,800…1,200 (adjustable via programming keys or analog input)
Signal processing
One-color-, Two-color-mode, Attenuation monitoring, Alarm, Average, Peak hold,
Valley hold, Advanced peak hold with threshold and hysteresis (adjustable via
programming keys or software)
1)
Measurements within specification over 5-95% of range
2)
at ambient temperature 23
5 °C
3)
= 1/ Response time 1 s
4)
with dynamic adaptation at low signal levels