36
-
S OFF
Activating (
ON
) and Deactivating (
OFF
) of the
Sighting Laser
. By pressing
Up
or
Down
the laser can be switched on and off.
SLOPE
The
Slope
is the quotient of t
he emissivity’s of both of the overlapping wavelengths and
therewith the deciding parameter for measurements in 2-color-mode.
EMISS 1.000
Setup of
Emissivity
. Pressing
Up
increases the value,
Down
decreases the value (also
valid for all further functions). The emissivity is a material constant factor to describe the
ability of the body to emit infrared energy
.
AVG 0.020
Setup of
Average time
. In this mode an arithmetic algorithm will be performed to
smoothen the signal. The set time is the time constant. This function can be combined
with all other post processing functions. The shortest value is 0,001 s. If the value is set to
0.0
the function is deactivated.
HOLD
Setup of
signal processing
. By pressing
Up
or
Down
the mode can be selected.
PEAK:
Setup of
Peak hold
. In this mode the sensor is waiting for descending signals. If
the signal descends the algorithm maintains the previous signal peak for the specified
time.
After the hold time the signal will drop down to the second highest value or will descend
by 1/8 of the difference between the previous peak and the minimum value during the
hold time. This value will be held again for the specified time. After this the signal will drop
down with slow time constant and will follow the current object temperature.
If the value is set to
0.0
the display will show
---
(function deactivated).
VALL:
Setup of
Valley hold
. In this mode the sensor waits for ascending signals. The
definition of the algorithm is according to the peak hold algorithm (inverted).
If the value is set to
0.0
the function deactivated.
APEAK
(Advanced Peak Hold): In this mode the sensor waits for local peak values. Peak