14
-
The temperature distribution in
the face is in this example
between 31 °C and 35.1 °C
(
). The size of the
canthus
is
approximately
assumed to be 15x15 mm.
Figure 8:
Temperature distribution face
To achieve the minimum temperature accuracy requirement, there must be at least 3x3 pixels within the
Canthus region (15x15 mm).
The following tables show the DOF ranges and the number of pixels in the canthus for the different available
optics depending on the distance.
Summary of Contents for BR 20AR
Page 1: ...Operators Manual BR 20AR Ambient temperature reference source...
Page 8: ...8 Figure 1 Dimensions mm ceiling mounting...
Page 9: ...Technical Data 9 Figure 2 Dimensions mm wall mounting...
Page 11: ...Installation 11 Figure 4 System overview with PI 450i camera...
Page 28: ...BR 20AR MA E2020 07 A...