LTDVE8CH– INSTRUCTIONS MANUAL
57
Bit 10: set if LD6 output stage has been disabled, reset if LD6 output stage is active
Bit 11: set if LD7 output stage has been disabled, reset if LD7 output stage is active
Bit 12: set if LD8 output stage has been disabled, reset if LD8 output stage is active
When any of the four thermal sensors on the output stages reports an excessive temperature, the
relevant error bit (1, 2, 3, or 4) is set and the relevant couple of output stages is shut down. This
happens when the temperature goes above about 90 °C. The same couple of output stages is then
reactivated when the temperature goes below about 80 °C.
When a hardware problem is detected on any of the eight output channels, the relevant error bit (5, 6,
7, 8, 9, 10, 11, 12 or 13) is set and the relevant output stage is permanently disabled. The error
condition may be reset by cycling the power supply off and on or by issuing a reboot command to the
controller (see
chapter 13.2.45
more information).
When the supply of the power stages is missing, the ERR LED is lit with a stable red colour. When
an over temperature condition or a hardware problem is detected the ERR LED is lit with a blinking
red colour.
Synchronization outputs SH1, SH2, SH3, SH4, SH5, SH6, SH7 and SH8 are never disabled.
Bit field [15:13] of this register is unused.
14.2.47. Registers CALIB_CUR_ADD[0:7]
These registers contain calibration constants for measurement of light output currents.
CALIB_CUR_ADD0
: calibration constant for measurement of light output LD1 current
CALIB_CUR_ADD1
: calibration constant for measurement of light output LD2 current
CALIB_CUR_ADD2
: calibration constant for measurement of light output LD3 current
CALIB_CUR_ADD3
: calibration constant for measurement of light output LD4 current
CALIB_CUR_ADD4
: calibration constant for measurement of light output LD5 current
CALIB_CUR_ADD5
: calibration constant for measurement of light output LD6 current
CALIB_CUR_ADD6
: calibration constant for measurement of light output LD7 current
CALIB_CUR_ADD7
: calibration constant for measurement of light output LD8 current
14.2.48. Registers CALIB_CUR_MUL[0:7]
These registers contain calibration constants for measurement of light output currents.
CALIB_CUR_MUL0
: calibration constant for measurement of light output LD1 current
CALIB_CUR_MUL1
: calibration constant for measurement of light output LD2 current
CALIB_CUR_MUL2
: calibration constant for measurement of light output LD3 current
CALIB_CUR_MUL3
: calibration constant for measurement of light output LD4 current
CALIB_CUR_MUL4
: calibration constant for measurement of light output LD5 current
CALIB_CUR_MUL5
: calibration constant for measurement of light output LD6 current
CALIB_CUR_MUL6
: calibration constant for measurement of light output LD7 current
CALIB_CUR_MUL7
: calibration constant for measurement of light output LD8 current