LTDVE8CH– INSTRUCTIONS MANUAL
42
FILTER_SEL0
: filter setting for input TR1
FILTER_SEL1
: filter setting for input TR2
FILTER_SEL2
: filter setting for input TR3
FILTER_SEL3
: filter setting for input TR4
FILTER_SEL4
: filter setting for input TR5
FILTER_SEL5
: filter setting for input TR6
FILTER_SEL6
: filter setting for input TR7
FILTER_SEL7
: filter setting for input TR8
Allowed values are in the range from 0 to 6 and are listed below. Avoid operation with non-listed
values.
When 0x0 filter is disabled (pass through) (default value)
When 0x1 filter is enabled with a 10 µs time constant
When 0x2 filter is enabled with a 20 µs time constant
When 0x3 filter is enabled with a 50 µs time constant
When 0x4 filter is enabled with a 100 µs time constant
When 0x5 filter is enabled with a 200 µs time constant
When 0x6 filter is enabled with a 500 µs time constant
Bit fields [15:3] of these registers are unused. When writing these bits, they must be set to zero.
14.2.8. Registers INPUT_SEL[0:15]
Each bit field [8:0] of these sixteen registers is the selector of the relevant input multiplexer. Each
input multiplexer feeds the dedicated pulse generator.
INPUT_SEL0
: setting of input multiplexer 0
INPUT_SEL1
: setting of input multiplexer 1
INPUT_SEL2
: setting of input multiplexer 2
INPUT_SEL3
: setting of input multiplexer 3
INPUT_SEL4
: setting of input multiplexer 4
INPUT_SEL5
: setting of input multiplexer 5
INPUT_SEL6
: setting of input multiplexer 6
INPUT_SEL7
: setting of input multiplexer 7
INPUT_SEL8
: setting of input multiplexer 8
INPUT_SEL9
: setting of input multiplexer 9
INPUT_SEL10
: setting of input multiplexer 10
INPUT_SEL11
: setting of input multiplexer 11
INPUT_SEL12
: setting of input multiplexer 12