3.Installation And Carry
3.1
Installation Process
3.2
Carry
Microscope is precision instrument, careful in carry. Shut down power supply
,
pull out power
line. Lock the eyepiece tube and condenser etc, no slice left on stage. Don't move
nosepiece , focusing knob, mechanical stage and eyepiece tube etc, don't make the
eyepiece off. Avoid stumble into the chair, violent shocks and collision can cause damage.
a. Single hand carry way: with one hand holding the body, the microscope will be hold by
thumb buttoned filed from inside.
b. Two hands lift way: one hand turn back the microscope slightly by holding body
handle ,the another hand holding the microscope’s forepart.
4. take out socket head wrench
for use whenever appropriate.
5. take out main body
,
loosen locking screw
,
put the seidentopf binocular(trinocular)
head aside.
6. take out midbody
,
align connector to main body
according to fixed direction
,
locking fastening screw.
7. plug seidentopf binocular(trinocular) eyepiece tube in midbody ,lock it with socket
head wrench.
8.
Pull down dust cover
,
insert two eyepieces in tube
,
check if in good position.
9.
Pull down nosepiece’s dust cover
,
take out objective and install in sequence for
use.
10.
Rotate Coarse focusing knob
,
lower rotating stage, screwing 10Xobjective in
objective hole which on the nosepiece’s outer ring with yellow mark
,
then switching
clockwise the nosepiece
,
in proper order to install objective.
11.
Place filter
、
compensator and nosepiece centering wrench on side for use.
12.
Check instrument if in good condition, switch on and turn on the power , following
operation method to check it if in normal operating condition..
Installation complete!
▲
During installing eyepiece, Please screw eyepiece and wrench nosepiece’s
cover ,not wrench eyepiece directly.
▲
When leaving factory, objective hole with yellow mark has be regarded as basic
hole, centering wrench hole has been sealed ,should install 10X objective (yellow
mark line).
▲
Please properly dispose packing waste after installation complete.
▲
Microscope is precise instrument, so installation should be careful in case
scratching optical lens.
▲
Please don’t touch eyepiece or optical lens of condenser with finger, otherwise will
effect normal operation.
1. Confirm the service environment is good for use.
2. Clean working desk
,
put microscope parts in the left side.
3. Dispose the packing waste properly
,
don’t left plastic bag or foam on the working
desk.