background image

 

-

6. Troubleshooting 

I

n the period of using Smart series microscope, if there is any trouble occurs, please 

referring to the following sheet listed some common troubleshooting resolve them or keep 
in touch with us or local 

agency

Trouble 

Causation 

Remedy 

Switch on but bulb dark 

No bulb 

Install bulb 

Plug is unreliable 

Check joint again 

Bulb is broken 

Replace bulb 

Fuse is broken 

Replace fuse 

Bulb is flickering or 
brightness is unsteady. 

Bulb is unstable 

Insert again 

Bulb is broken 

Replace bulb 

Brightness of view field 
isn’t enough or is uneven. 

Bulb specification doesn’t meet the 
requirement. 

Replace bulb 

Bulb brightness is low. 

Rotate potentiometer to adjusting 
brightness.

 

Objective isn’t in correct optical path. 

Rotate the objective in correct 
position. 

The size of iris aperture is too Small. 

Adjust the size of iris aperture. 

Lens (objective, eyepiece, 
condenser, light collector) has dust. 

Clean it 

Position of Condenser is too low . 

Adjust condenser properly. 

Image    isn’t    clear 
(contrast    or definition 
isn’t    enough) 

Cover glass of specimen doesn’t 
meet the requirement. 

Use required thickness cover 
glass (0.17mm). 

Cover glass of specimen isn’t in up 
direction. 

Place specimen correctly. 

Surface of objective lens Is dirty 
(especially it is easy for the front lens 
of 4

  objective to dip in immersion 

oil). 

Clean it 

Immersion oil isn’t used for   
100× objective (oil) 

Use immersion oil 

Immersion oil doesn’t meet the 
requirement. 

Use immersion oil supplied by us. 

There is bubble in immersion oil. 

Clear the bubble away 

Size of iris aperture isn’t proper. 

Adjust the size of iris aperture. 

Position of condenser is too low. 

Readjust again 

One side of image is 
dark or image is 
moving as focusing. 

Objective isn’t in correct 
optical path. 

Make the objective in correct 
position.. 

Specimen isn’t placed correctly. 

Place specimen levelly on stage 
and clip it with clamp. 

Dirt and dust can be seen 
in Field of view. 

Dirt or dust on bulb glass 

Clean it   

Dirt or dust on specimen 
Condenser’s front lens has dust or 
dirt. 
Lens (objective, eyepiece, 
condenser, light collector) has dust. 

Objective touching 
specimen as changing low 
power to high power 

Cover glass of specimen 
doesn’t meet the requirement. 

Place specimen correctly. 

Cover glass is too thick. 

Use required thickness cover 
glass (0.17mm). 

Image observed by two 
eyes aren’t in 
superposition entirely. 

Interpupilary distance isn’t adjusted 
correctly. 

Adjust Interpupilary distance. 

Diopter isn’t adjusted correctly 

Adjust diopter 

Left and right eyepiece is different. 

Replace same eyepieces. 

It is easy for eyes to be 
tired during observing. 

Interpupilary distance isn’t adjusted 
correctly. 

Adjust Interpupilary distance. 

Diopter isn’t adjusted correctly. 

Adjust diopter 

Brightness isn’t enough 

Adjust brightness 

 

Summary of Contents for A15.2603

Page 1: ...A15 2603 Polarizing Microscope Instruction Manual To ensure the safety and obtain satisfactory performance please study this instruction manual thoroughly be fore start to use the instrument...

Page 2: ...CONTENT Attentions 2 1 Parts Name 3 2 Performance Parameter 4 3 Installation And Carry 5 4 Operation And Use 6 5 Maintenance 7 6 Troubleshooting 9 Appendix outfits 10...

Page 3: ...e alcohol gasoline paper and other combustibles near the instrument to prevent a fire 5 Notes on replacing the bulb The instrument used with 12V 20W or 6V 30W halogen lamp of G4 LED model with 3 4V 3W...

Page 4: ...g Knobs Brightness control knob 4 Compensator Built in Polarizer Clamps Analyzer driving lever Stand Bertrand Lens Nosepiece Center Adjustable rotating Stage Diopter Adjusting Ring Midbody Power Fuse...

Page 5: ...ar Trinocular head Interpupillary Distance 48 mm 75 mm Dioper Adjusting 5 With antifungus device 3 Midbody Analyser 360 Division 2 30 With locking device Bertrand lens with center Adjustable Combining...

Page 6: ...down nosepiece s dust cover take out objective and install in sequence for use 10 Rotate Coarse focusing knob lower rotating stage screwing 10Xobjective in objective hole which on the nosepiece s out...

Page 7: ...ion 3 gypsum slip first class red Orthogonal polarizing appear first class purplish red interference colours 4 Mica 4 slip Orthogonal polarizing appear first class grey colours 5 Quartz wedge Orthogon...

Page 8: ...ement of halogen Please use specified bulb in case of damage or fire Specified bulb halogen 12V 20W or 6V 30W LED 3 4V 3W LED According to outfits 1 shut off the power switch to side unplug the power...

Page 9: ...in fuse tube socket again 5 5 Not in use 1 Please shut off the power covered by dust cover keep in dry and cool place 2 The eyepiece and objective should be placed in dry containers such as moisturep...

Page 10: ...objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use immersion oil Immersion oil doesn t meet the requirement Use immersion oil supplied by us There is bubble...

Reviews: