Revolving nosepiece not in click stop 
position 

Revolve to click-stop position 

Image tinged yellow 

Blue filter not used 

Use blue filter 

Insufficient illumination brightness  Condenser aperture too small 

Readjust aperture 

Condenser position too low 

Correct position 

Dirt  or  dust  on  the  lens  (condenser, 
objective, eyepiece, collector) 

Cleaning 

 
2. Mechanical 

Trouble   

Cause 

Remedy 

Image  not  focusable  with  high 
power objective 

Slide upside-down 

Turn slide cover 

Cover glass too thick 

Use cover glass of specified thickness 
(0.17mm) 

High power objective contacts slide 
when  changed-over  from  low 
power 

Side upside down   

Turn slide over 

Cover glass to thick 

Use cover glass of specified thickness 
(0.17mm) 

Specimen movement unsmooth 

Mechanical  stage  not  securely 
fastened 

Tighten all fastener 

Binocular images not integrated 

Interpupillary  distance  not  correctly 
adjusted 

Adjustment 

Excessive eye fatigue 

Incorrect diopter adjustment 

Correct adjustment 

Inadequate brightness or illumination    Adjust brightness with control dial 

                 
3. Electrical 

Trouble 

Cause 

Remedy 

Lamp dose not light when switched 
ON 

No electrical power   

Check power cord connection 

Lamp bulb not inserted 

Insert correctly 

Lamp failure 

Replacement 

Sudden lamp failure 

Unspecified  lamp  bulb  used  or  input 
voltage too high 

Replace  with  specified  lamp  bulb.  If 
the  same  symptom  occurs  after 
replacing  the  lamp,  contact  your 
dealer 

Insufficient illumination brightness  Unspecified lamp bulb used 

Replace with specified lamp bulb 

Voltage too low 

Increase brightness with control dial 

Flickering 

or 

unstable 

lamp 

brightness 

Lamp bulb about to fail 

Replacement 

Lamp bulb not correctly inserted into 
socket 

Check for positive connection 

 
7. Maintenance 
1. Cleaning Lenses 
Dust is best removed with a soft brush or gauze. 
More persistent dirt, such as fingerprints, grease and oil, may be removed with soft cotton, lens tissue, or 
gauze lightly moistened with absolute alcohol (ethyl or methyl alcohol). 
To clean immersion oil off the oil-immersion type objective, use lens tissue, soft cotton or gauze lighty 
moistened with petroleum benzine only. 

Summary of Contents for A12.1010

Page 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Page 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Page 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Page 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Page 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Page 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Page 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Page 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Page 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Page 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Page 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Page 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Page 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Page 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Page 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...

Page 16: ...er thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surface...