    5.4 Focus by 10X objective 
          Put the 10X objective into the optic path and focus by turning coarse and fine focusing knob 
         

Turn the stage and focusing knob according to the direction shown in the picture 

         

There is no coarse focusing knob in one side of stage’s moving knob and in the other side there are 
both coarse and fine focusing knob 

         

Never turn the left and right focusing knob in the adverse direction at the same time. Never turn 
the coarse focusing knob when the stage is in the end of removing range. Otherwise it will make 
damage to the instrument 

         

 It is difficult to do focusing if turning the focusing knob optionally. The specimen may be crushed 

when 100X objective is used. Please read the instruction carefully to avoid making damage to the 
cover glass or objective 

                5.4.1 Put the 10X or 4X objective into optic path 
                5.4.2 Remove the stage to the top by turning the coarse 
  focusing knob 
                5.4.3 Observe through the eyepiece and turn the coarse 
  focusing knob slowly to let the stage down and down till there is   
sharp image 
                5.4.4 Turn the fine focusing knob to focus finely 
                          Use the 10X objective or 4X objective firstly before 
  40X and 100X objective. Turn the fine focusing knob to focus finely 
         
 
 
 
 
    5.5 Adjusting the diopter for eyepieces 
          Adjust the diopter ring according to user’s diopter for left and   
right eyes. This function could make use of advantages of objectives   
fully. Meanwhile it also can be react as focusing 
          5.5.1 Put the 40X objective in the optic path and turn the   
coarse and fine focusing knob to do focusing 
          5.5.2 Put the 10X objective (or 4X objective) in the optic path 
          5.5.3 Let your left eye observe through the left eyepiece and do   
focusing by the diopter adjusting ring in left eyepiece 
          5.5.4 Repeat the above steps till sharp image can be observed   
by left and right eyepiece at the same time. 
 
    5.6 Adjusting the upright position for condenser 
          Turn the lifting knob for condenser till it reaches the top position then fall it a little bit down. If there 

is dispersion image in the view field, please remove the condenser a little bit of up and down to make 
the dispersion image disappeared 

 
    5.7 Objectives 
          Turn  the  nosepiece  to  choose  the  objective  and  adjust  the  view  field  diaphragm  and  aperture 

Summary of Contents for A12.1010

Page 1: ...A12 1010 Laboratory Biological Microscope Instruction Manual Please read this manual carefully before use microscope...

Page 2: ...Contents BM2000 1 Structure and nomenclature 4 2 Technical parameter and specification 6 3 Configuration 8 4 Assemblage 9 5 Operation 13 6 Troubleshooting 18 7 Maintenance 20 8 Products Standards 20...

Page 3: ...oscope on a stable plane and keep it in balance Keep the environment breezy and cover the microscope with the dust cover 2 6 Do not place microscope in moist room to avoid short circuit Please turn of...

Page 4: ...inserted into the holder firmly Such Operation may cause short circuit or smoking At last please make sure if the collector has been mounted properly 9 Temperature for illumination There will be high...

Page 5: ...re screw Nosepiece Objectives Specimen holder Mechanical stage Condenser with auxiliary lens Collector Binocular head Eye tubes fixture Microscope body Power supply Lightness adjusting knob switch scr...

Page 6: ...Working environment Temperature 0 40 Humidity max 85 no dew Altitude within 2000m Pollution 2 Assembly 2 Optical parameter 2 1 Objectives infinity achromatic objectives Magnification N A Cover glass t...

Page 7: ...tem Content Piece 1 Microscope s body including stage nosepiece coaxial coarse and fine focusing adjustment bracket for condenser adjusting power supply halogen bulb and normal collector 1pc 2 Siedent...

Page 8: ...ded AC connector and make sure that the electronic cord has been connected safely Please use the supplied electronic cord by manufacturer Please make sure to use PE electronic cord when using extra lo...

Page 9: ...rsion image disappeared 4 5 2 Objectives The objectives have been mounted in the microscope before shipment The objectives have been mounted in the microscope before shipment Take off the specimen fro...

Page 10: ...lacement for fuse 4 6 2 1 Turn off the power supply turn the switch to O and pull out the connector 4 6 2 2 Open the cover for fuse by screwdriver 4 6 2 3 Mount the new fuse 4 6 2 4 Make sure that the...

Page 11: ...there is sharp image 5 4 4 Turn the fine focusing knob to focus finely Use the 10X objective or 4X objective firstly before 40X and 100X objective Turn the fine focusing knob to focus finely 5 5 Adjus...

Page 12: ...ctive when the aperture diaphragm adjusting shaft reaches some position If the objective has been changed please move the shaft to the position with the same value which is in the objective in the opt...

Page 13: ...4 2 Never press the bottle overly as it may caused oil gushed out of the bottle 5 9 4 3 Clean the out surface of the bottle from remanent oil 5 10 Adjusting the coarse focusing knob s tension The tens...

Page 14: ...or dust on the slide Cleaning Condenser position too low Correct position Poor image quality low resolution poor contrast No cover glass on the specimen Attach cover glass Cover glass too thick or thi...

Page 15: ...Inadequate brightness or illumination Adjust brightness with control dial 3 Electrical Trouble Cause Remedy Lamp dose not light when switched ON No electrical power Check power cord connection Lamp b...

Page 16: ...er thinner ect to clean the painted or plastic surfaces of the instrument We recommend the use of silicon cloth More persistent dirt may be cleaned with mild detergent solution Printed plastic surface...

Reviews: