background image

 
 
 

 

A12.0911

 

Biological Microscope 
 

Instruction Manual 

 
 

 
 
 
 

This instruction manual is for the operation guide, troubleshooting and 
maintenance  to  the  biological microscope.  Please  study  this  manual 
thoroughly before operating, and keep it with the instrument. The 
manufacturer reserves the rights to the modifications by technology 
development.  On  the  basis  of  operation  ensured,  technical  specifications 
may  be  subject  to  changes  without  notice. 

Summary of Contents for A12.0911

Page 1: ...g and maintenance to the biological microscope Please study this manual thoroughly before operating and keep it with the instrument The manufacturer reserves the rights to the modifications by technol...

Page 2: ...Adjust the Field Diaphragm 10 3 11 Aperture Diaphragm 11 3 12 Use the Oil Objective 100X 11 3 13 Use the Filter 12 3 14 Use Button and Interface of the Digital Head 12 4 Assembling and Operation of Ac...

Page 3: ...replacing the LED lamp and wait until the lamp and base cool down completely see Fig 3 Bulb selected only single 3W LED class 3B 5 External power supply adaptor is adopted Wide voltage range is suppor...

Page 4: ...on t place these chemical near to fire or fire source For example when turning on or turning off the electrical device please use these chemical in a ventilated place 2 Don t use organic solution to w...

Page 5: ...minator Light Adjustment Knob Coarse Focusing Knob Fine Focusing Knob Focus Arm Koehler Illuminator Condenser Stage Tension Adjustment Ring Objective Body Group Eyepiece Binocular Head Objective Nosep...

Page 6: ...Illuminator Condenser Koehler Illuminator Color Temperature Knob Power Switch LCD Display Coarse Focusing Knob Fine Focusing Knob Focus Arm Stage Objective Objective Nosepiece Body Group Gemel Binocul...

Page 7: ...Koehler Illuminator Condenser Focus Arm Fine Focusing Knob Koehler Illuminator LCD Display Power Switch Transmitted Color Temperature Knob Coarse Focusing Knob series biological microscope with LED fl...

Page 8: ...epiece Body Group Stage Focus Arm External Power Socket Coarse Focusing Knob Light Adjustment Knob Transmitted Fine Focusing Knob Koehler Illuminator Eyepiece Gemel Binocular Head LED Fluorescence Ill...

Page 9: ...Following is the Assembling Scheme and the numbers denote the assembling order Before assembling make sure there is no dust dirt or other materials which will disturb it Assemble carefully and do not...

Page 10: ...20X 40X 100X objective matching with the nosepiece hole 1 2 3 4 5 see Fig 6 When operating first use the low magnification objective 4X or 10X to search for specimen and focus and then replace with hi...

Page 11: ...Power Cord Don t use strong force when the power cord is bended or twisted otherwise it will be damaged 1 Make sure the power switch is at O OFF before connecting the power cord 2 Insert the connector...

Page 12: ...ording to the demand Insert a flat tip spanner into the color temperature adjustment knob to select increase or decrease with step 100 Rotate clockwise to increase color temperature and rotate counter...

Page 13: ...nob for clear details and lock the position screw The position screw can stop the objective touching the clips 3 6 Adjust the Focusing Tension If the handle is very heavy when focusing or the specimen...

Page 14: ...o raise it to the highest position see Fig 19 2 Rotate the objective 10X to the light path and focus the specimen 3 Rotate the field diaphragm adjustment ring to put the field diaphragm to the smalles...

Page 15: ...ective 3 12 Use the Oil Objective 100X 1 Use the 4X or 10X objective to focus the specimen 2 Place a drop of oil on the specimen observed see Fig 23 3 Rotate the nosepiece counterclockwise and rotate...

Page 16: ...e indicator flashes to indicate that the camera is working 2 Photo button when the camera is working press once to take a photo 3 Recording button when the camera is working the first press is the rec...

Page 17: ...nto the power socket of the microscope back and then insert the other end into the power supply socket and ensure them be connected well see Fig 28 Before connecting the external power adaptor firstly...

Page 18: ...necting the power cable firstly rotate the rechargeable battery box switch the transmission power switch and the reflection adjustment knob with switch to Off turn off 2 Turn on the battery box switch...

Page 19: ...C UL safety certification of PC 5 For binocular observation after the image is clear view the output video through the corresponding video capture analysis software If the image is unclear rotate the...

Page 20: ...ntered it doesn t need to be adjusted in operation in normal cases Fig 37 4 4 2 Assembling and Operation of Disc Phase Contrast Condenser 1 Assembling of Condenser please refer to 2 2 1 change the obj...

Page 21: ...denser flapper socket as the direction of the arrow pointed see Fig 39 2 Each diaphragm or hole has corresponding position when it sounds ka da in dark field flapper rotating it indicates that one dia...

Page 22: ...to the USB output port on the back of observation head and the other end into the USB interface of PC see Fig 41 View the microscope video by the corresponding video capture and analysis software and...

Page 23: ...lean the specimen Stains have accumulated on the lens Clean the lens 4 Unclear image The cover glass is not standard Use a standard cover glass with thickness 0 or 0 17mm The cover glass faces down Ad...

Page 24: ...e 4 Stage declines itself cannot stay on the focal plane when observing Tension knob is too loose Tighten a little 5 Coarse focusing knob can t decline The base of the condenser is too low Raise the b...

Page 25: ...field is dark The fluorescence filter block adjustment knob is at Brightfield Rotate the fluorescence filter block adjustment knob to Fluorescence 3 The background is yellowgreen when in brightfield o...

Reviews: