Optima OP-W Series User Manual Download Page 5

Contents

Scale main unit

Scale pan

User Manual

AC adapter : 12V

Preparation

1. Carefully remove contents from the package which may include fragile parts.
2. Carefully assemble the scale by placing the scale pan.
3. Place the scale in a firm and flat surface, and avoid location where the scale 

maybe subject to vibration, sunlight, airflow, contact with water, and strong 
electromagnetic wave. 

4. Level the scale by using the bubble level indicator (if available) and adjusting the 

feet of the scale.

5. Locate scale in the following condition:

1. Temperature : 0

o

C ~ 40

o

C

6. Plug in the scale to a surge protected power outlet if using the AC adapter.  You 

may use the internal rechargeable batteries also once it has charged.

7. Press the ON/OFF key in the front panel to turn the scale on and wait for it to 

finish initialization (i.e. when the display shows a weight of 0).

8. Let the scale warm up for 1 minute before calibration and use.
9. Make sure any weight placed on the scale is not beyond the rated maximum.
10.If scale is not used for an extended period of time, turn off the scale and unplug 

the AC adapter (if it’s connected).

 

OP-W Series

 

User Manual / Getting Started

 

5/25

Summary of Contents for OP-W Series

Page 1: ...OP W Series User Manual...

Page 2: ...void overloading or dropping material onto the platform as that could damage the balance Material that has a static charge could influence the performance of the balance Electromagnetic interference c...

Page 3: ...nu 6 Units Menu 8 Si CAL 01 Menu 11 CALIB Menu 15 FILTER Menu 17 OIML Menu 17 QUIT Menu 20 RESET Menu 20 C1 C2 Menu 21 ADC Menu 21 RST Menu 22 ZERO Menu 22 G1SET Menu 23 G2SET Menu 24 More Info 25 Tro...

Page 4: ...g Optima scales for your important application Here is the list of models this manual covers Model Max Weight grams Accuracy grams OPW P3 3 000 0 2 OPW P6 6 000 0 5 OPW P15 15 000 1 OPW P30 30 000 2 O...

Page 5: ...the scale 5 Locate scale in the following condition 1 Temperature 0oC 40oC 6 Plug in the scale to a surge protected power outlet if using the AC adapter You may use the internal rechargeable batterie...

Page 6: ...r the scale has been turned on and in the normal weighing mode to get to the settings menu follow the procedures below Press the PLU key holding it for more than 2 seconds Press ENTER key OP W Series...

Page 7: ...guration the user can change For each settings press Enter key to change the value of that particular parameters by using the up down keys to increase decrease the parameters Press Enter key to confir...

Page 8: ...r will go back to normal weighing mode Units Menu Select the weighing unit the place of the decimal the maximum capcacity and the first rang division and the second rang division if it exists Select t...

Page 9: ...se the number of decimals after the decimal point Press Enter key On the following screen enter the maximum capacity of the scale For the example below a 5kg max is set Note the decimal places that wa...

Page 10: ...rang or to exit from the Units menu If the unit does not have the second rang option the message will not show the configuration of the second rang is the same as the configuration of the first rang T...

Page 11: ...ogrammed from the setting menu G1set and G2set explained later in the document For example we will calibrate the scale in 3 steps 0kg 0 5kg and 1kg When the pan is empty press Enter key The screen wil...

Page 12: ...ammed by using the arrow keys Put the mass of 0 5kg on the pan let it stabilize and press Enter key The screen will blink while taking the value of the weight When the measurement is done Si 02 will b...

Page 13: ...using the arrow keys Put the mass of 1kg on the pan let it stabilize and press Enter key The screen will blink while measuring the weight value When the measurement is done Si 03 will be shown to indi...

Page 14: ...tings menu press the left arrow key To exit with saving configuration select the menu Quit and press Enter key To continue with more configurations using the left and right arrows to traverse and pres...

Page 15: ...n in 2 steps Press the Enter key and the internal counts is displayed Note may not be exactly the same as this number Press Enter key When the pan is empty press the Enter key to calibration the zero...

Page 16: ...to program the value of 1kg use the arrow keys to change to 1kg Put the calibration weight in this case 1kg on the pan and press Enter key The screen will blink while taking the value and the setting...

Page 17: ...NOT follow OIML recommendations 1 Scale does follow OIML recommendations There are 3 sub parameters RS Manual zero limit Adjusts 0 99 from the bottom of the scale It s the capacity for zero ing when...

Page 18: ...be programmed Press Enter key With this configuration of the RS sub parameter the scale is capable of absorbing the zero deviation 2 from the bottom of the scale For example if the bottom of the scale...

Page 19: ...is configuration of the OT sub parameter the scale can absorb the zero drifting which are inferior or equal to 40 of the division value For example if the scale is configured for the division of 0 001...

Page 20: ...at the menu Quit as shown below To quit configuration menu without saving changes press U esc key RESET Menu The RESET menu resets all the configuration to the default settings Note that this may lea...

Page 21: ...working simultaneously ADC Menu This menu is composed of two sub menus SR speed of ADC 5 10 20 20 is the highest speed GN Amplification of load cell 1 2 2 is when increase of amplification is needed...

Page 22: ...press U esc key ZERO Menu This menu allows adjusting the number of divisions starting from which it will return to zero There can be 6 values with 0 meaning that the function is not acting The values...

Page 23: ...acceleration value without decimal point of the location you are interested in If this value is modified a linear calibration Menu Si CAL 01 needs to be performed Press Enter key To quit with save se...

Page 24: ...librated without the need to re calibrate Press Enter key Enter the gravity acceleration value without the decimal point of the new location s earth gravity Press Enter key To quit with save select Qu...

Page 25: ...xternal interference Please revisit the Preparation section of this manual Also please make sure the calibration procedure has been performed correctly It won t hurt to run through the calibration pro...

Reviews: