With this configuration of the OR sub-parameter, the scale is capable of staring with
an initial deviation of the load cell by 16% from the bottom of the scale.
Press “Enter” key.
With this configuration of the OT sub-parameter, the scale can absorb the zero
drifting which are inferior or equal to 40% of the division value.
For example, if the scale is configured for the division of 0.001 kg = 1 g, 0.4 g zero
deviations will be absorbed.
Press “Enter” key.
To exit and save the configuration, select “Quit” menu and press “Enter” key.
OP-W Series
User Manual / Scale Settings