![Optima OP-W Series User Manual Download Page 12](http://html1.mh-extra.com/html/optima/op-w-series/op-w-series_user-manual_1642636012.webp)
Press the “Enter” key to program the calibration value.
In the example above, the value 0.5kg has been programmed by using the arrow
keys.
Put the mass of 0.5kg on the pan, let it stabilize, and press “Enter” key.
The screen will blink while taking the value of the weight.
When the measurement is done, “Si-02” will be shown to indicate that it’s ready for
a 3rd measurement.