Si.CAL-01 Menu
This setting enables user to make a linear calibration.
Calibration can be done with 2 steps up to 10 steps maximum.
Take the Earth Gravity into consideration. The gravity parameters can be
programmed from the setting menu “G1set” and “G2set” (explained later in the
document)
For example, we will calibrate the scale in 3 steps: 0kg, 0.5kg, and 1kg.
When the pan is empty press “Enter” key.
The screen will the blinking while measuring zero.
“Si-01” is shown to indicate that the measure of zero has been taken and that the
unit is ready to take another measurement.
OP-W Series
User Manual / Scale Settings