Page 50
12. Troubleshooting
Review the information in the table below to troubleshoot operating problems.
PROBLEM
CAUSE
SOLUTION
I. Optical Section:
The illumination is ON, but the field of
view is dark.
The plug is not connected to the illu
-
mination
Connect the cable
The brightness is too low
Adjust to a proper setting
The edge of the field of view is vignetted
or the brightness is asymmetric.
The incident illuminator is not correctly
oriented
Change the angle of the incident illu
-
minator
Dust and stains can be seen in the field
of view.
There are stains and dust on the spec
-
imen
Clean the specimen
There are stains and dust on the eye
-
piece
Clean the eyepiece
Poor image quality:
• The image is not sharp
• The contrast is not high
•
The details are not clear
•
Image glares
The lenses (additional lens, objective,
eyepieces) are dirt
Throughly clean all the optical system
One side of the image is out of focus. The specimen is out of place (tilted)
Place the specimen flat on the stage.
II. Mechanical Section:
The focus knob is hard to turn.
The tension adjustment collar is too
tight
Loosen the tension adjustment collar
The focus is unstable.
The tension adjustment collar is too
loose
Tighten the tension adjustment collar
III. Electric Section:
The LED doesn’t turn on.
No power supply
Check the power cord connection
The brightness is not enough
The brightness adjustment is low
Adjust the brightness
The light blinks
The power cord is poorly connected
Check the power cord
IV. Viewing tube assembly:
The field of view of the two eyes is dif
-
ferent
The interpupillary distance is not cor
-
rect
Adjust the interpupillary distance
The diopter correction is not right
Adjust the diopter correction
The viewing technique is not correct,
and the operator is straining the eye
-
sight
When look into the eyepieces, do not
stare at the specimen but look at the
whole field of view. Periodically, move
the eyes away to look at a distant ob
-
ject, then back into the eyepieces
V. Microphotography and video:
The image is unfocused
Incorrect focusing
Adjusting the focus system as in the
present manual
Bright patches appear on the image
Stray light is entering the microscope
through the eyepieces and through the
camera viewfinder
Cover the eyepieces and the viewfind
-
er with a dark cloth
Summary of Contents for SZ-STL1
Page 15: ...Page 15 7 Instrument description 7 1 System Diagram...
Page 67: ...Pagina 67 7 Descrizione dello strumento 7 1 Diagramma di Sistema...
Page 119: ...P gina 119 7 Descripci n del instrumento 7 1 Diagrama del Sistema...
Page 171: ...Page 171 7 Description de l instrument 7 1 Diagramme de Syst me...
Page 223: ...Seite 223 7 Beschreibung des Instruments 7 1 Systemdiagramm...
Page 275: ...P gina 275 7 Descri o do instrumento 7 1 Diagrama de Sistema...