Page 48
10. Microphotography
①
②
①
②
F
Fig. 80
ig. 80
F
Fig. 81
ig. 81
F
Fig. 82
ig. 82
③
④
①
②
③
④
10.1 Installing the C-mount adapter
1.
Loosen the clamping screw
①
on the trinocular port and re
-
move the dust cap
②
. (Fig. 80)
2.
Screw the C-mount adapter
③
to the camera
④
and insert
the round dovetail of the C-mount into the empty hole of the
trinocular port (Fig. 81), then tighten the clamping screw
①
.
10.2 Use of reflex cameras
1.
Insert the Reflex adapter
①
into the relay tube
②
.
2.
Screw the “T2” ring
③
(not provided) to the reflex adapter.
3.
Connect the Reflex camera
④
to the “T2” ring just installed.
(Fig. 82)
4.
Mount the other end of the relay tube
②
into the empty hole
of the trinocular port, then tighten the clamping screw. (Fig.
80)
•
“T2” ring is not provided along with the microscope, but is
commercially available.
•
While shooting dark specimens, darken eyepieces and view
-
finder with a dark cloth to minimize the diffused light.
•
To calculate the magnification of the camera: objective mag
-
nification * camera magnification * lens magnification.
• When using an SLR camera, mirror movement may
cause the camera to vibrate.
•
We suggest lifting the mirror, using long exposure times
and a remote cord.
Summary of Contents for SZ-STL1
Page 15: ...Page 15 7 Instrument description 7 1 System Diagram...
Page 67: ...Pagina 67 7 Descrizione dello strumento 7 1 Diagramma di Sistema...
Page 119: ...P gina 119 7 Descripci n del instrumento 7 1 Diagrama del Sistema...
Page 171: ...Page 171 7 Description de l instrument 7 1 Diagramme de Syst me...
Page 223: ...Seite 223 7 Beschreibung des Instruments 7 1 Systemdiagramm...
Page 275: ...P gina 275 7 Descri o do instrumento 7 1 Diagrama de Sistema...