OPTEX FA CS-30TAC Instruction Manual Download Page 14

14

Setup of Functions

 AADJ

   OFF

 AADJ

     ON

       L

 EXIt

   100

    120

   120

   SET

       H

 EXIt

       L

   SET

       H

   SET

   250

   300

   300

   SET

2 Point Adjustment

(1) Aim the sensor head to the target. Confirm the target is sufficiently 

larger than the field of view.

(2) In the Setting Operation select the Adjustment Setting, and then 

press Enter button.

(3) Select the 2 Point Adjustment, and then press the ENTER button.

(4) Select ON, and then press the ENTER button.

(5) Select SET, and then press the ENTER button.

* To set the temperature H (higher value) first, select EXIT.

(6) The current measurement value is displayed.

 

Input the temperature L (lower value) of the target, and then press 

the ENTER button.

This setting is for adjusting the measurement value with the specified value 

according to the measuring targets. A intended value can be output by setting 

L (lower value) and H (higher value) conforming to the both specified value.

* Measurement accuracy can not be guaranteed for the value made by 2 

Point Adjustment.

* Do not change the emissivity rate after setting in 2 Point Adjustment, 

otherwise the adjusted value will be changed.

(7) Aim the sensor head to the target. Confirm the target is sufficiently 

larger than the field of view.

(8) Select SET, and then press the ENTER button.

* To input the temperature H (higher value) later, select EXIT.

(9) The current measurement value is displayed.

 

Input the temperature H (higher value) of the target, and then 

press the ENTER button.

Make sure to perform Save Settings.

* If the timings to set temperature L and H are different (e.g., when using the same target for 

setting L and H), set either of them first and perform Save Settings. Otherwise the 2 Point 

Adjustment will be cancelled.

* The adjusted value are effective after the both of temperature L and H are saved. 

While 2 Point Adjustment is on, 2 Point Adjustment indicator is 

displayed on sub display.

If the setting of temperature L or H value is not saved, the 

measurement value is output based on the setting stored previously. 

But 2 Point Adjustment indicator is displayed on the sub display.

EXIT

   300

  AADJ

Adjust the measurement values (2 points) to the temperature values (2 points) of the target. 

Summary of Contents for CS-30TAC

Page 1: ...S 30TAC CS 40TAC CS 30TAC HT CS 40TAC HT 91 Chudoji Awata cho Shimogyo ku Kyoto 600 8815 JAPAN TEL 81 75 325 1314 FAX 81 75 325 2936 OPTEX FA CO LTD Non Contact Thermometer Printed in JAPAN 0807230 20...

Page 2: ...to convert the infrared energy emitted from the surface of an object into temperature This thermometer measures the surface temperature of solid and liquid without contacting them The temperature of...

Page 3: ...off DO NOT TOUCH THE LENS Do not touch the lens with hard or sharp objects Do not insert foreign objects into the light receiving part Otherwise incorrect measurement will occur KEEP THE THERMOMETER A...

Page 4: ...Emissivity rate adjustment 0 1 to 1 2 Power supply 12 to 24 VDC 10 Consumption 120 mA Max load 80 mA Eco mode Ambient temperature Sensor head 0 to 180 C Amplifier 0 to 65 C Sensor head 0 to 100 C Amp...

Page 5: ...ens surface 7 M12 P1 0 10 30 14 Sensor head Sensor cable 4 3000 MAX Cable cover Serial No buttons ENT button Output cable 6 2000 MAX Cable No Main display Amplifier Sub display 64 52 35 39 31 mm DIN r...

Page 6: ...2 Do not wire 15 Orange Black 2 Connect to GND 16 4 Shielded cable Output cable 1 Connect to the 4 20 mA input of an analog device Analog output allowable load 250 and analog output impedance 47 2 Ex...

Page 7: ...Insert the connector to the cable cover The metal plate of connector should be placed onto the clasp in the cable cover Sensor cable 5 15 5 Cable tie of approx 2 mm width Guide wire 1 2 3 4 mm Connect...

Page 8: ...al screw is M12 P1 0 Fix securely into the hole of 12 mm or more using the attached hexagon nut The optional mounting fitting can help you adjust the angle easily Mounting Mount the sensor head perpen...

Page 9: ...w shown above When measuring a high temperature target keep as much distance from it as possible within the range of the field of view Otherwise it may cause incorrect measurement When using the optio...

Page 10: ...ation Press 3 sec or more Setting Operation Make sure to perform Save Settings If you cancel in Save Settings all settings entered are not saved 8 8 8 8 8 8 8 8 3 To confirm or change the settings pre...

Page 11: ...d alarm ON OFF can be set The Unit has 3 Alarm Output Modes as below NORMAL Mode Alarm output is kept on as long as the value is above or below the limit value DELAY Mode Alarm output turns on when a...

Page 12: ...e bank 1 to 4 to change the setting Settings of Response Time Adjustment Analog Output Alarm Output Trigger Input and Display Off can be saved in each bank 1 In the Setting Operation select the Bank S...

Page 13: ...ently larger than the field of view 2 In the Setting Operation select the Adjustment Setting and then press Enter button 3 Select the TEACH Adjustment and then press the ENTER button 4 The current mea...

Page 14: ...not change the emissivity rate after setting in 2 Point Adjustment otherwise the adjusted value will be changed 7 Aim the sensor head to the target Confirm the target is sufficiently larger than the f...

Page 15: ...40 L 0 3 L Lower limit value is displayed Input the temperature L lower limit value and then press Enter button Make sure to perform Save Settings The minimum width of output range is 100 degrees 0 t...

Page 16: ...on when the value exceeds or falls below the limit value and is kept for a specific time NORMAL Mode Alarm output is kept on as long as the value is above or below the limit value DELAY Mode Alarm ou...

Page 17: ...value and then press the ENTER button 6 Select NORMAL Mode and then press the ENTER button Make sure to perform Save Settings 6 Select DELAY Mode and then press the ENTER button 7 The DELAY Time is d...

Page 18: ...og output types as below HOLD MIN MIN HOLD P P P P HOLD SAMP SAMPLE Setup of Functions MIN HOLD The minimum value between the triggers is output HOLD MAX MAX P P HOLD The P P value maximum minus minim...

Page 19: ...hen press Enter button Make sure to perform Save Settings WAVE Mode HOLD MAX HOLD MIN WIMT 200 WLIM 150 WIMT 200 WLIM 150 Analog Output is controlled by External Trigger Analog Output is controlled by...

Page 20: ...ing to the analog output of MAX HOLD set by the trigger input alarm is output while the value is outside the range between H and L ECO OFF 1 In the Setting Operation select the Display Off Setting and...

Page 21: ...gger mode MAX MIN MAX MIN MAX MIN MAX MIN Trigger temperature Display Off Time to turn off s s s s OFF Note of the settings Save or cancel the settings entered in the Setting Operation Switch other ba...

Page 22: ...s is dirty remove the dust using a blower for cleaning lens For stubborn dirt apply a small amount of ethyl alcohol to a cotton swab or special lens cleaning cloth and gently wipe off the dirt Amplifi...

Reviews: