11
Function List
To change the settings, press the ENTER button for three seconds or more.
The settable items are shown below.
Change the settings as necessary.
While the indication flashes, the settings are being read or written and the button operation
is not available.
No.
Function
Descriptions
➊
Bank
Setting
Select the bank (1 to 4) to change the setting.
Settings of Response Time, Adjustment, Analog Output, Alarm Output, Trigger Input, and
Display Off can be saved in each bank. Select the bank by cable connections.
➋
Response
Time
Setting
Select the response time.
You can select 1 to 200 of the moving average times.
The larger number, the longer response time and the smoother fluctuation.
➌
Adjustment
Setting
Adjust the measurement value.
The Unit has 3 methods for the adjustment as below;
TEACH Adjustment
: Adjust the measurement value to the temperature value of the target.
Emissivity Rate Adjustment
: Input the emissivity rate directly.
2 Point Adjustment
: Adjust the measurement values (2 points) to the tempera
-
ture values (2 points) of the target.
➍
Analog
Output
Setting
Adjust the analog output (4-20 mA) range.
You can setup higher limit value (value for 20 mA) and lower limit value (value for 4 mA).
➎
Alarm
Output
Setting
Make the alarm output settings.
Alarm Output Mode, temperature of alarm output and alarm ON/OFF can be set.
The Unit has 3 Alarm Output Modes as below;
NORMAL Mode
: Alarm output is kept on as long as the value is above or below the limit value.
DELAY Mode
: Alarm output turns on when a specific time period passes after
the value exceeds or falls below the limit value.
ONE-SHOT Mode
: Alarm output turns on when the value exceeds or falls below the
limit value and is kept for a specific time.
➏
Trigger
Input
Setting
Make the trigger input settings.
Trigger input mode, analog output type, and trigger ON/OFF can be set.
The Unit has 2 Trigger Modes as below;
External Mode
: Analog Output is controlled by External Trigger.
WAVE Mode
: Analog Output is controlled by the judgment criterion value (WAVE LIMIT).
The Unit has 4 analog output types as below;
MAX HOLD
: The maximum value between the triggers is output.
MIN HOLD
: The minimum value between the triggers is output.
P-P HOLD
: The difference between the maximum and minimum values between the
triggers is output. (For External Mode only)
SAMPLE HOLD
: The value at the moment of the trigger is output. (For External Mode only)
➐
Display Off
Setting
Make the display off settings.
You can reduce consumption current by turning off the display during Normal Operation.
If the unit is left unoperated for 60 seconds or more, it returns to Normal Operation. In this
case, the settings are cancelled. Make sure to perform Save Settings.
MODE
8888
END
SET
END
SET
BANK
1
Flashing / Button operation is not available
Lighting / Button operation is available