Handset
3.
Bed Operation and Maintenance
The motorised bed functions can be operated via the handset.
To avoid damage, the handset should always be hung up (e.g. on the mattress
base or side rail) when not in use.
3.4
Using the Handset
Whilst operating the bed, the rear castors of the bed must be braked to
ensure stability during bed adjustments. The Opera Solo Bed should only
be moved to reposition the bed.
On the handset the up arrow on the controller indicates raising of the
corresponding part of the bed. The down arrow indicates lowering of the
corresponding part of thebed.
The bed is equipped with anti-entrapment features. If the bed platform is
higher than 320mm from the floor, long press the bed platform down button
to lower it to below 320mm. To continue lowering the bed, release and long
press the button again. Ensure that there are no limbs, objects or obstacles
within the beds moving range.
Where a connection fault is identified and rectified, the control system
will require a reset in order to operate again. To reset the control system,
press and hold down the back rest and knee break buttons (up and down)
simultaneously for 6 seconds. If a back up battery is fitted it will beep
intermittently during the system reset.
Backrest up/down
Knee break up/down
Backrest and Knee
break up/down
Bed platform up/
down
Reverse Trendelenburg
action/return to normal
level