Unpack the contents
Unpack the contents of the Signature Dual Link Kit and place to one side.
If the beds have side skirts, remove the skirts from the inner side of each
bed where they will be joined together. Refer to the Signature Profiling Bed
Installation Guide on how to do this and the below step.
Unbrake the castors of the Signature Profiling Beds
and position the two
beds parallel to one another, with at least two meters between them. Brake
the castors once positioned.
Using the bed handset, electronically adjust the backrest of the beds to
gain access to the control boxes, this is where the link kit will be connected
to in the next steps.
1
Only unbrake the castors when the beds need moving to prevent rolling away and
injury. Turn off all power to the bed(s) after adjusting.
Signature bed
Two parallel Signature beds
2.
Installation and Commissioning
Read before you begin!
Read this Installation Guide conscientiously before
installing the Signature Dual Link Kit
Signature bed with back and leg rest raised