Page 2: ...Care Signature Dual Link Kit Installation Guide Version 1 1 Explanations of Symbols 2 2 Installation and Commissioning 3 3 General Information 11 4 Guarantee 13 5 Opera Warranty Terms and Conditions 13 Contents ...
Page 3: ...ions This information is safety relevant This product must be disposed of in a seperate refuse collection in the European Union Do not dispose of as normal domestic waste Read Instructions i This symbol indicates hazards due to electrical voltage There is mortal danger ELECTRICAL 1 Explanations of Symbols ...
Page 4: ...ink Kit Installation Guide Version 1 A Receiver x 1 B Controller x 1 C Y Splitter Cable x 1 D Cable Clip x 4 F U Clamp x 2 E Actuator to Receiver Cable x 1 G Cable Connection Cover x 3 2 1 Contents 2 Installation and Commissioning ...
Page 5: ... one another with at least two meters between them Brake the castors once positioned Using the bed handset electronically adjust the backrest of the beds to gain access to the control boxes this is where the link kit will be connected to in the next steps 1 Only unbrake the castors when the beds need moving to prevent rolling away and injury Turn off all power to the bed s after adjusting Signatur...
Page 6: ...s per photos below Connect the two pronged plug of the Y splitter cable 3 to the power cable 1 Connect the socket on the splitter cable 4 to the two pronged head of the short power supply cable 2 Take a connection cover G and push the joined cables the plug and socket into the cover this should cover where the two cables connect 2 1 2 F 3 4 2 4 2 2 4 3 1 G C 2 Installation and Commissioning ...
Page 7: ...o pronged head 6 and connect it to the remaining socket on the Y splitter cable C Place the joined cable into a connection cover G 3 Connect the Receiver On the actuator to receiver cable E take the round plug socket 8 and connect to the plug 9 on the receiver A 4 6 7 6 8 8 9 E C G E E A 2 Installation and Commissioning ...
Page 8: ...ent actuator cable from the control box this is the actuator plug that is colour coded with a yellow sticker both on the control box and on the plug Connect the receiver to actuator cable E socket 10 to the two pronged plug of the height adjustment actuator cable 11 5 10 11 10 E E 2 Installation and Commissioning ...
Page 9: ... cable as per the diagram below Press the edge of the clips down to lock the clips Secure the receiver Take two cable clips D and attach them to the inside of the mattress platform side channel this needs to be at the side where the two beds will be joined together Place the cable clips approx 5cm apart and approx 13cm from the chassis securing bolt 12 With one hand pull open the security clip and...
Page 10: ...air the controller Press the up and down buttons of the controller simultaneously until the LED on the receiver starts blinking The blinking confirms successful pairing To confirm successful pairing use the controller see photos below to raise and lower the bed by pressing the up and down buttons The distance between the receiver and controller should be no less than two meters during synchronisat...
Page 11: ... bed with the controller both beds should simultaneously raise and lower together Attach the beds together Unbrake the castors on the beds and push the two frames together so that they are flush Take one U bracket F and place it over the two joined frames at the head end of the bed Insert the U bracket bolts through the holes in the bottom of the bracket and tighten with an Allen key to secure Pla...
Page 12: ...ignature Dual Link Kit for a use not outlined in this instruction manual Modifying the Signature Dual Link kit 3 3 Warnings Before installing the Signature Dual Link Kit it is recommended that the Signature Dual Beds do not have any anything attached or on them this includes Grab rails side rails and mattresses During the briefing specific attention must be drawn to any potential dangers for the f...
Page 13: ...ad the Signature Profiling Bed Installation Guide and Technical Specifications V2 before installing the Signature Dual Link Kit and operating the beds Failure to do so may result in serious injury entrapment and electric shock When operating the adjustment functions of the bed ensure that the cables do not get caught or pulled out 3 4 Service Life and Disposal The service life is dependent upon th...
Page 14: ...he product 5 1 2 This contractual product warranty does not operate to limit rights under the statutory warranties referred to in clause 5 3 1 below 5 2 Warranty Conditions 5 2 1 Proof of purchase invoice must be provided when requesting service under warranty 5 2 2 Opera requires any customer requesting service under the warranty to comply with directions from Opera staff in relation to troublesh...
Page 15: ...he product the customer will be provided with credit for Opera products or may be refunded the price of the product at Operas election This credit or refund will be for the amount of the purchase price of the product excluding the associated delivery cost 5 2 6 In the event that a replacement refund or store credit is provided as per section 5 2 5 the faulty item will become the property of Opera ...
Page 16: ...derstorm activity force majeure voltage supply problems tampering or unauthorised repairs by an persons use of defective or incompatible accessories exposure to abnormally corrosive conditions or entry by any insect vermin or foreign object in the product d To damage arising during transportation installation or while moving the product or to any transportation costs of the product or any parts th...
Page 17: ...oss of use loss of profits or revenue or for any resulting indirect or consequential loss of damage b Opera s aggregate liability in respect of all claims under the warranty shall not exceed the original purchase price of the product or at Opera s option the replacement of the product with a like or similar product c Opera excludes all other warranties conditions terms representations and undertak...
Page 18: ...17 Opera Care Signature Dual Link Kit Installation Guide Version 1 Notes ...
Page 19: ...18 opera care co uk 0333 222 8584 Notes ...
Page 20: ...Opera Care Azure House Connaught Road Kingswood Hull HU7 3AP 0333 222 8584 support opera care co uk opera care co uk Version 1 Published September 2020 ...