![Opera Signature Dual Link Kit Installation Manual And Technical Specifications Download Page 11](http://html1.mh-extra.com/html/opera/signature-dual-link-kit/signature-dual-link-kit_installation-manual-and-technical-specifications_1640181011.webp)
10
opera-care.co.uk
|
0333 222 8584
|
9
Connect the second bed
On the second bed, unplug the height adjustment actuator cable from the
control box. This is the actuator colour coded with a yellow sticker.
Take the actuator to receiver cable (E) and connect it to the height
adjustment actuator, make sure that the long actuator to receiver cable is
under the mattress platform and out of the way of any moving parts.
Check that the second bed is connected to the receiver by raising and
then lowering the bed with the controller, both beds should simultaneously
raise and lower together.
Attach the beds together
Unbrake the castors on the beds and push the two frames together so that
they are flush.
Take one U bracket (F) and place it over the two joined frames at the head
end of the bed.
Insert the U bracket bolts through the holes in the bottom of the bracket
and tighten with an Allen key to secure.
Place the second U bracket over the two frames at the footend of the bed
and secure with the U bracket bolts.
10
ATTENTION: Use both of the supplied U brackets to ensure that the beds are
secured together. Always brake the castors after the beds have been repositioned.
2.
Installation and Commissioning