4003_5001_D02
Page 14 of 35
4.5.5.
DATE
Selecting this item allows the date to be adjusted. Ensure that a valid date is entered in
DD/MMM/YY format. Refer to Section 4.12 RTC BACKUP.
4.5.6.
TIME
Selecting this item allows the time to be adjusted. Ensure that a valid time is entered in 24hour,
HH:MM format. Refer to Section 4.12 RTC BACKUP.
4.5.7.
THRESHOLD
CAUTION: Adjusting this parameter without first consulting Online Electronics Ltd may cause
the unit to miss pig passages or give false detections.
Selecting this item allows the THRESHOLD level to be configured to control the sensitivity of the
unit. The standard THRESHOLD value is 35mG. Reducing the THRESHOLD value increases the
sensitivity. The THRESHOLD can be configured to any value between 5mG and 9999mG.
4.5.8.
POWER MODE
Selecting this item allows the power mode to be configured. In STANDARD POWER MODE the
display will be ON all of the time. The display draws significant power from the batteries which
will reduce the battery life of the unit. In LOW POWER MODE the display will turn OFF after 10
seconds to preserve the battery life. The detection routines are identical in each mode.
4.5.9.
STABILISATION DELAY
Selecting this item allows the STABILISATION delay to be configured. It is possible to configure
any value between 00:00:00 and 99:59:59 in HH:MM:SS format. Refer to Section 4.2
STABILISATION for further information on the STABILISATION delay functionality.
4.5.10.
DETECTION DELAY
Selecting this item allows the DETECTION delay to be configured. It is possible to configure any
value between 00:00:00 and 99:59:59 in HH:MM:SS format. See section 4.4 DETECTION MODE
for further information on the DETECTION delay functionality.
4.5.11.
DETECTION METHOD
Selecting this item allows the detection method to be configured. The options available are:
PEAK
–
50%
PEAK
–
25%
PEAK
–
10%
PEAK
–
05%
DETECTION ONLY
PEAK
–
xx% determine the PIG PASSED detection criteria, reference 4.4.1 PIG PASSED, and the
DETECTION ONLY selects the pig detection only detection criteria, reference section 4.4.2 PIG
DETECTION ONLY.