4003_5001_D02
Page 27 of 35
Examples of strings received:
4003,01234,M,00045,012,EXT,-48,7<CR>
Indicating 4003 serial number 1234, magnitude output, peak flux at detection of 45mG, event
number 12, externally powered, Rssi of -48dBm and SNR of 7.
4003,00345,N,00125,018,4.3,-40,6<CR>
Indicating 4003 serial number 345, north pole detected, peak flux at detection of 125mG, event
number 18, battery voltage of 4003 is 4.3V, Rssi of -40dBm and SNR of 6.
4003,00346,S,00101,001,4.2,-48,7<CR>
Indicating 4003 serial number 346, south pole detected, peak flux at detection of 101mG, event
number 1, battery voltage of 4003 is 4.2V, Rssi of -48dBm and SNR of 7.
An optional ‘verbose mode’ is availab
le on the Online Electronics LoRa Base Station which outputs
a serial string with the same information but in a human readable format. The final
“
ACK sent
”
string indicates that the Online Electronics LoRa Base Station sent an acknowledgement to the
the 4003 unit sending the message.
Pig Detect 01234
North 02258
Event 017
External Supply
RssiValue=-37 dBm, SnrValue=7
ACK sent