Taking a test
Checking if the result is in range
Example range
OneTouch Verio® Level 3
Control Solution Control
Range 102-138 mg/dL
Each vial of test strips
has both OneTouch Verio®
Level 3 Control Solution
and OneTouch Verio®
Level 4 Control Solution
ranges printed on its
label. Compare the result
displayed on the meter
to either the OneTouch
Verio® Level 3 Control
Solution or OneTouch
Verio® Level 4 Control
Solution range printed
on the test strip vial
,
depending on the type of
control solution you used.
CAUTION:
The range printed on the test strip vial is for control
solution tests only
and is not a recommended range for
your blood glucose level.
OneTouch Verio® Level 4
Control Solution Control
Range 298-403 mg/dL