27
ERROR MESSAGES AND DETAILS ABOUT YOUR SYSTEM
Detailed information about your system
Comparing meter and laboratory results
Test results with the OneTouch® UltraMini® Meter are plasma-calibrated. This helps you and your healthcare professional to compare your
meter results with laboratory tests. If you have been using another type of meter—one that provides whole blood-calibrated results—you
may notice that your test results with the OneTouch® UltraMini® Meter are approximately 12% higher. OneTouch® UltraMini® Meter test results
and laboratory test results both are expressed in plasma-equivalent units. However, your meter result may differ from your laboratory result
due to normal variation. Meter results can be affected by factors and conditions that do not affect laboratory results in the same way.
Your OneTouch® UltraMini® Meter glucose value is considered accurate when it is within ±20% of the laboratory measurement. There are some
specific situations that could cause a difference of more than ±20%:
• You have eaten recently. The blood glucose level from a fingertip can be up to 70 mg/dL higher than blood drawn from a vein
(venous sample) used for a lab test.
1
• Your hematocrit (percentage of blood that is red blood cells) is high (above 55%) or low (below 30%).
• You are suffering from severe dehydration.
• You tested at a temperature near the low end of the operating range (43°F) and you get a high glucose result (i.e., greater than
180 mg/dL). In this situation, repeat the test in a warmer environment with a new test strip as soon as possible.
For accuracy and precision data and for important information on limitations, see the insert that comes with your test strips. To maximize your
chances of an accurate comparison between meter and laboratory results, follow a few basic guidelines:
Before going to the lab
• Perform a control solution test to make sure the meter is
working properly.
•
Do Not
eat for at least eight hours before you test your blood.
• Take your meter with you to the lab.
While at the lab
• Conduct your meter test within 15 minutes of the lab test.
• Use only fresh, capillary blood obtained from the fingertip.
• Follow all instructions in this User Guide for performing a
blood glucose test with your meter.
1. Sacks, D.B.: “Carbohydrates.” Burtis, C.A., and Ashwood, E.R. (ed.),
Tietz Textbook of Clinical Chemistry
. Philadelphia: W.B. Saunders Company (1994), 959.
Symbols
Cautions and Warnings: Refer to safety-related notes in the User Guide and inserts that came with your meter and testing supplies.
Direct current
Low battery
SN
Serial Number
Electrical and safety standards
The meter has been tested for immunity to Level 4 electrostatic discharge as specified in IEC 61000-4-2. This meter has been tested for
immunity to radio frequency interference over the frequency range 80 MHz to 2.5 GHz at 3 V/m as specified in IEC 61000-4-3. Degree of
protection rating: Minimum of IP2X. This meter complies with CISPR 11:2003, Class B (Radiated Only). Emissions of the energy used are low and
are not likely to cause interference in nearby electronic equipment.
C US
CAN/CSA C22.2 61010-1:04, UL 61010-1:04, IEC 61010-1 and IEC 61010-2-101.
Guarantee
LifeScan guarantees that the OneTouch® UltraMini® Meter will be free of defects in material and workmanship for three years, valid from the
date of purchase. The guarantee extends only to the original purchaser and is not transferable.
Technical specifications
Reported result range
20–600 mg/dL
Calibration
Plasma-equivalent
Sample
Fresh capillary whole blood
Test time
5 seconds
Assay method
Glucose oxidase biosensor
Meter power source
One replaceable 3.0 Volt CR 2032
lithium battery (or equivalent)
Unit of measure
mg/dL
Memory
500 blood glucose test results
Automatic shutoff
2 minutes after last action
Size
4.25 x 1.26 x .67 inches
Weight
Approximately 1.4 ounces, with battery
Operating ranges
Temperature: 43–111
°
F
Altitude: up to 10,000 feet
Relative humidity: 10–90%
Hematocrit: 30–55%
Battery ratings
One 3.0 V d.c., 3 mA
(one CR 2032 battery)
direct current