Standard Specifications
ZR-RX40 Start Up Guide
3
SPECIFICA
T
IONS
43
Analog
input
section
Measurement
accuracy
(*2) (*3)
Resistance
temperature
detector
Type
Measurement
Temperature Range (
°
C)
Applied current
Measurement
Accuracy
Pt100
–200 to 850 (FS = 1050
°
C) 1 mA
±1.0
°
C
JPt100
–200 to 500 (FS = 700
°
C) 1 mA
±0.8
°
C
Pt1000
–200 to 500 (FS = 700
°
C) 0.2 mA
±0.8
°
C
Maximum input voltage
60 Vp-p (b/– terminals, between input terminals, between
input terminal/GND)
Reference contact compensation
Internal/External switching
Input impedance
1 M
Ω
±5 %
Allowable signal source resistance
300
Ω
or less
Temperature coefficient
Gain: 0.01 % of F.S./
°
C
Withstand voltage
350 Vp-p (between input channel/GND; between each channel) 1
minute
Insulation resistance
Between input terminal/GND: At least 50 M
Ω
(at 500 VDC)
External
Input/
Output
Sections
Logic input/
Pulse input
Number of channels
(*4)
4 ch
Pulse
input
modes
Revolutions
mode
Spans
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M RPM/F.S.
Maximum number of pulse
inputs
50 k/sec
Counts
mode
Spans
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M C/F.S.
Maximum number of pulse
inputs
50 k/sampling interval
Inst. Mode Spans
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M C/F.S.
Maximum number of pulse
inputs
50 k/sampling interval
Trigger input
Number of channels
1 ch
Specifications
of each input
section
Maximum input
voltage
24 V
Threshold voltage
Approx. 2.5 V
Hysteresis range
Approx. 0.5 V (2.5 to 3 V)
Alarm output
Number of channels
4 ch
Output format
Open collector output (pull-up resistance: 10 k
Ω
)
5 to 24 VDC, up to 100 mA
Output conditions
Level judgment, window judgment, logic pattern judgment, pulse
judgment
Main unit specifications
ZR-RX40A