7
5. Specifications
Product description
Automatic Blood Pressure Monitor
Model
HEM-7156
Display
LCD digital display
Cuff pressure range
0 to 299 mmHg
Pulse measurement range 40 to 180 beats / min.
Blood pressure measurement range SYS: 60 to 260 mmHg / DIA: 40 to 215 mmHg
Accuracy
Pressure: ±3 mmHg / Pulse: ±5% of display reading
Inflation
Automatic by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
Operating mode
Continuous operation
IP classification
Monitor: IP20 / Optional AC adapter: IP22 for HHP-OH01, IP21 for other AC adapters
Rating
DC6 V 4.0 W
Applied part
Type BF (arm cuff)
Power source
4 “AA” batteries 1.5 V or optional AC adapter (INPUT AC 100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Battery life
Approximately 1,000 measurements (using new alkaline batteries)
Durable period (Service life)
Monitor: 5 years / Cuff: 5 years / Optional AC adapter: 5 years
Operating conditions
+10 to +40 °C / 15 to 90% RH (non-condensing) / 800 to 1060 hPa
Storage / Transport conditions
-20 to +60 °C / 10 to 90% RH (non-condensing)
Contents
Monitor, arm cuff, 4 “AA” batteries, storage case, Instruction Manual and
Protection against electric shock
Internally powered ME equipment (when using only batteries)
Class II ME equipment (optional AC adapter)
Weight (not including batteries)
Monitor: approximately 340 g / Arm cuff: approximately 163 g
Dimensions (approximately value) Monitor: 105 mm (W)× 85 mm (H)× 145 mm (L)
Arm cuff: 145 mm × 532 mm (air tube: 750 mm)
Memory
Stores up to 60 readings
Note
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of ISO 81060-2:2013. In the clinical validation study, K5 was used on 85
subjects for determination of diastolic blood pressure.
• IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are protected
against solid foreign objects of 12.5 mm diameter and greater such as a finger. The optional AC adapter (other than HHP-OH01) is protected
against vertically falling water drops which may cause issues during a normal operation. HHP-OH01 is protected against oblique falling water
drops which may cause issues during a normal operation.