![background image](http://html1.mh-extra.com/html/omron/hbp-1320/hbp-1320_instruction-manual_741557037.webp)
35
EN
Immunity test
EN60601-1-2
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
80% AM (1 kHz)
3 V/m
80 MHz to
2.7 GHz
80% AM (1 kHz)
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the HBP-1320, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommend separation distance
d
= 2
150 kHz to 80 MHz
d
= 2
80 MHz to 2.7 GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to he
transmitter manufacturer and
d
is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters as
determined by an electromagnetic site
survey
a
, should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the HBP-1320 is used exceeds the applicable RF
compliance level above, the HBP-1320 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the HBP-1320.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
2829490-0B_IM-HBP-1320-E-03-04-2020.book Page 35 Thursday, April 1, 2021 8:24 AM
Summary of Contents for HBP-1320
Page 39: ...37 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 37 Thursday April 1 2021 8 24 AM ...
Page 79: ...77 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 77 Thursday April 1 2021 8 24 AM ...
Page 119: ...117 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 117 Thursday April 1 2021 8 24 AM ...
Page 159: ...157 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 157 Thursday April 1 2021 8 24 AM ...
Page 199: ...197 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 197 Thursday April 1 2021 8 24 AM ...
Page 239: ...237 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 237 Thursday April 1 2021 8 24 AM ...
Page 281: ...279 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 279 Thursday April 1 2021 8 24 AM ...
Page 321: ...319 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 319 Thursday April 1 2021 8 24 AM ...
Page 323: ......
Page 324: ......
Page 326: ...۳۷ ...