TE CONNECTIVITY SENSORS   

/// USER NOTICE M210_EN                              03/2018

 

Page 1 

 

 

 

 

 

 

 

 

 
 
 
 
 
Important note 

 

The 

M210

 device is an electronic which puts forward its analogue output before its display (the display works as an 

output copy). All metrological check has to be done thru the analogue output. 

 

1. Before you start 

 

 

Connect the sensor to the conditioning module as per the wiring scheme. 

 

Ensure that the wires are correctly tightened into the terminal block. 

 

Connect the 

M210

 to the 230 Vac – 50Hz supply or DC supply (depending on version). 

 

Wait a few moments for the display to stabilise (warm-up time). 

 

NOTE:

  The  standard  M210  can  function  with  115  V 

–  50  Hz  (main  supplied  version)  by  changing  an  internal 

jumper. 

 

 

 

 

 

USER NOTICE M210

 

Summary of Contents for M210

Page 1: ...the analogue output 1 Before you start Connect the sensor to the conditioning module as per the wiring scheme Ensure that the wires are correctly tightened into the terminal block Connect the M210 to...

Page 2: ...hind the front panel on the bottom right hand side above the zero potentiometer The calibration function simulates a pressure or force as determined during calibration To simulate the pressure or forc...

Page 3: ...ter The gain potentiometer is hidden behind the front panel on the bottom left hand side above the zero potentiometer The calibration function simulates a pressure or force as determined during calibr...

Page 4: ...o 28 V 0 25 A Operating temperature range 0 to 50 C RH 95 at 40 C no condensation Storage temperature range 40 to 85 C Dimensions w x h x d 96 x 48 x 155 mm Drilling dimensions 92 x 45 mm Mass 500 gra...

Page 5: ...dc 5 mA short circuit protected Bandwidth at 3 dB 1000 Hz Amplification accuracy 0 01 Output impedance 10 Noise 10 mV pk to pk Sensitivity A O 20 potentiometer Zero adjustment A O 20 potentiometer 5 D...

Page 6: ...display To define the decimal point position on the display points 1 2 or 3 on the left of the display are connected with a solder point see the drawing below Short circuit None for 1999 1 for 1 999 2...

Page 7: ...a included in this document RG1 GAIN RESISTANCE CALCULATION FORMULA 7 8 10 435 1 4 Ve RG RG1 is in k Ve corresponds to the sensor sensitivity in mV for Full Scale Example Pressure transducer 1 bar Sen...

Page 8: ...e the values of the analogue outputs and the display when the test pushbutton is pressed these two values correspond to the calibration values To calibrate the measurement chain once the pushbutton is...

Page 9: ...INAL SUPPLY 1 Earth 2 Phase or V 3 Phase or V 10 PIN TERMINAL MEASUREMENT 1 Sensor shield 6 Current output Option I 2 Sensor signal 7 Current output Option I 3 Sensor signal 8 Voltage output 4 Sensor...

Page 10: ...USER NOTICE M210 TE CONNECTIVITY SENSORS USER NOTICE M210_EN 03 2018 Page 10...

Page 11: ...corresponding potentiometer to obtain the desired threshold level on the display Release the pushbutton The unit is ready to function Note No calibration pushbutton is present on the M210 AL1 or AL2 B...

Page 12: ...USER NOTICE M210_EN 03 2018 Page 12 C Six pin connector wiring scheme 1 Common LT 4 Common HT 2 Rest LT 5 Rest HT 3 Work LT 6 Work HT Rear Panel Front Panel Note SH High Threshold SB Low Threshold TH...

Page 13: ...se TE Connectivity s obligations shall only be as set forth in TE Connectivity s Standard Terms and Conditions of Sale for this product and in no case will TE Connectivity be liable for any incidental...

Reviews: