24
Maintenance
Maintenance Inspection and Safety Management
The HBP-1320 must be maintained to ensure functionality and to secure the safety of patients and operators.
Daily checks and maintenance should be performed by the operator. (page 25)
In addition, qualified personnel are necessary to maintain the performance and the safety, and to conduct
periodic inspections. We recommend that the verification test be performed at least once a year.
Cleaning of the Device
Cleaning should be performed in accordance with your facility’s practice.
Surface cleaning
Wipe with a cloth that has been moistened with isopropyl alcohol diluted to 50 v/v%, or ethyl alcohol
(disinfection alcohol) diluted to 80 v/v% or less and wrung out.
Do not wipe the DC jack or allow it to become wet.
Removing dust
Use a moistened cotton bud to remove dust that has accumulated on the vent ports.
Service
The device requires no routine service other than cleaning, and visually checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the manufacturer.
Cuff Care
Wipe clean on the surface of the cuff with a cloth moistened with a 70 v/v% dilution of isopropyl alcohol,
or a 80 v/v% or less dilution of disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the inside well.
Example of connection for pressure accuracy confirmation:
1. Display the pressure accuracy confirmation screen as explained in
“Utility Mode” on page 17.
2. Connect the blood pressure monitor, the calibrated reference pressure
gauge, and the cuff and inflation bulb.
3. Check the pressure value of the blood pressure monitor and the
pressure value of the calibrated reference pressure gauge.
Note:
• Make sure that the blood pressure monitor reading is within ±3 mmHg compared to the calibrated
reference pressure gauge.
• To rapidly deflate the cuff, press the [ ] button.
To repeat “Pressure Accuracy Confirmation”, turn off the power and repeat the procedure from
step 1 in “Utility Mode” on page 17.
Display the pressure value.
Display “0” in the pressure accuracy confirmation screen.
2829490-0B_IM-HBP-1320-E-03-04-2020.book Page 24 Thursday, April 1, 2021 8:24 AM
Summary of Contents for HBP-1320
Page 39: ...37 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 37 Thursday April 1 2021 8 24 AM ...
Page 79: ...77 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 77 Thursday April 1 2021 8 24 AM ...
Page 119: ...117 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 117 Thursday April 1 2021 8 24 AM ...
Page 159: ...157 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 157 Thursday April 1 2021 8 24 AM ...
Page 199: ...197 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 197 Thursday April 1 2021 8 24 AM ...
Page 239: ...237 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 237 Thursday April 1 2021 8 24 AM ...
Page 281: ...279 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 279 Thursday April 1 2021 8 24 AM ...
Page 321: ...319 2829490 0B_IM HBP 1320 E 03 04 2020 book Page 319 Thursday April 1 2021 8 24 AM ...
Page 323: ......
Page 324: ......
Page 326: ...۳۷ ...