After 6 seconds the Control Solution reading
will be displayed on the display screen.
Step 5:
Compare the reading on the screen to
the Control Solution Range on the bottle of test
strips.
If the number displayed on the meter is with in
the L1 number range printed on the Test Strip
Bottle, the Meter and Strips are working properly
and are ready for testing. If the reading falls
outside the Level of LOW (L1) range printed on the test strip
bottle (see page 28, “Control Solution Trouble Shooting”).
Follow steps 3 through 5 for the Level 2 Control Solution
test and compare the result with the Level 2 range printed
on the test strip bottle.
NOTE:
• Control Solution testing results will be stored in the Meter’s
memory and indicated by CTL icon. Control Solution testing result
will not be used for calculating averages.
• Remove the test strip from the meter into proper waste container.
In approximately 2 minutes the meter will automatically turn off.
IMPORTANT:
DO NOT REUSE TEST STRIPS.