10.3.7.1.
Sound: This button sets the volume of the tone when keys are pressed and the volume
of the empty tank alarm
10.3.7.2.
Language: This button cycles through the different languages provided on the
system.
10.3.7.3.
Timeout If the Laser system is in Ready mode but not being used, after a certain
amount of time the system will automatically revert to Standby mode. This is to ensure
that the Laser is not accidentally fired if the foot pedal is pressed inadvertently. The
timeout button will change this amount of time, and is shown in minutes.
10.3.7.4.
Date: This reflects the current date.
10.3.7.5.
Time: This allows the time of the system to be updated.
10.3.7.6.
Error Log: This function is used by service technicians.
10.3.7.7.
Advanced Settings: This is an advanced setting screen that is only accessible by
OmniGuide personnel.
10.3.8.
Volume
This button controls the volume of the firing noise.
10.3.9.
Ready/Standby Button
The ready/standby button controls when the Laser will be allowed to fire if the foot pedal is
pressed. This button is only active in the Laser output control screen, and is disabled for all other
screens.
10.4.
OmniGuide Fiber Instructions for Use
OmniGuide’s BeamPath™ flexible fiber connects to a CO
2
Laser via a custom optical
Adapter. The fiber has a hollow core through which CO
2
Laser energy and inert gas
travel. These flexible fibers can be used either with one of several surgical specific
handpieces, or through an endoscope, to facilitate access to difficult to reach anatomy.
The OmniGuide fiber transmits a CO
2
Laser beam, along with a flowing gas, to a
treatment site as directed by a physician.
Please read the instructions for use (IFU)
included with all fiber products to ensure proper usage.
_____________________________________________________________
If uncertain about any portion of this IFU, contact OmniGuide for assistance.
10.4.1.
Preoperative System Preparation
Visually inspect the fiber adapter and verify that there is no discernible damage, that
all parts are clean and there are no loose parts or loose connections.
Wipe the Adapter, gas input port, and ST connection with sterile gauze and sterile
saline. For sterile cases, use aseptic technique to attach the sterile Gas Filter Unit (P/N
ACC-GFU-100) to the output tube.