![OmniGuide FELS-25A User Manual Download Page 18](http://html1.mh-extra.com/html/omniguide/fels-25a/fels-25a_user-manual_739296018.webp)
L-1633 Rev. 3
© OmniGuide, Inc.
Page 18
5.1.5.
Gas Management Data Connection
If helium gas supplied from helium gas tanks on the laser system cart, the system has
the ability to display the amount of gas available in the gas tanks on the laser touch
screen. The gas management data connectors on the back of the laser and on the gas
management system have to be connected by a gas management data cable supplied
with the system (P/N 1-0142-014-00-01).
5.1.6.
Door Interlock
The door interlock connection allows the use of the door interlock safety feature, as
may be required at your health care facility. If the door to the room where the laser is
being used is opened, this feature will automatically disable operation of the laser. If
door interlock operation is not required, a door interlock plug, (supplied with the
system, P/N 1-0205-007-01-03), can be plugged into the door interlock connection at
the back of the laser to defeat the interlock and allow operation of the laser. The door
interlock feature is implemented through the laser software.
5.1.7.
RFID Antenna
RFID antenna is built into the fiber adapter and is a component of the system RFID
functionality that allows the laser system to recognize specific Omniguide fiber and
adjust accordingly the default settings and settings limits for the power and gas
pressure. The specific fiber type is indicated on the laser touch screen.
5.1.8.
OmniGuide Fiber Adapter
To produce sufficient output power at the distal end of an OmniGuide fiber, the laser
beam generated in the laser is delivered precisely to the input end of the fiber by means
of fiber adapter with a standard ST socket for fiber connection. The adapter has a
quick-disconnect gas port for the helium gas supply and incorporates an RFID antenna
as shown in figure 5.3.